Theory of fast capillary gas chromatography. Part 1: Column efficiency
Author:
- 1 November 1997
- journal article
- research article
- Published by Wiley in Journal of High Resolution Chromatography
- Vol. 20 (11) , 597-604
Abstract
Typically, fast GC requires high column pressure drop. Under that condition, known relations do not provide a complete description of dependence of [apparent] plate height, ĥ, on average velocity, u, of carrier gas.A complete descriptions for ĥ vs. u is derived and analyzed. It is shown that, under the high pressure drop, ĥ = B/u2 + C1 u2 + C2 u. This is substantially different from the Van Deemter equation ĥ = B/u + Cu for the low pressure drop. Analytical expressions for the minimum plate height and for the respective optimum u under the high pressure drop are derived and analyzed.Keywords
This publication has 8 references indexed in Scilit:
- Relations between pneumatic parameters of ideal gas in a column with uniform permeabilityChromatographia, 1995
- Constant flow versus constant pressure in a temperature‐programmed gas chromatographJournal of High Resolution Chromatography, 1995
- Capillary GC on 50 micrometer I.D. Columns coated with thick films. Theory and selected practical resultsJournal of High Resolution Chromatography, 1995
- Molecular basis of peak width in capillary gas chromatography under high column pressure dropAnalytical Chemistry, 1993
- Increased speed of analysis in directly coupled gas chromatography-mass spectrometry systemsJournal of Chromatography A, 1981
- Plate Height in Gas ChromatographyAnalytical Chemistry, 1960
- Influence of Pressure Gradients on Resolution in Gas ChromatographyAnalytical Chemistry, 1959
- Longitudinal diffusion and resistance to mass transfer as causes of nonideality in chromatographyChemical Engineering Science, 1956