Effect of breath holding on blood flow measurement using fast velocity encoded cine MRI

Authors:

Abstract
Breath‐hold MR measurement of cardiac output was compared with results from respiratory triggered MR acquisitions, since flow measurement during breath‐holding may be different from physiological blood flow. Cardiac output during large lung volume breath‐holding (4.47 ± 0.63 l/min in the aorta and 4.53 ± 0.59 l/min in the pulmonary artery) was significantly lower than that measured during normal breathing (6.09 ± 0.49 l/min and 6.48 ± 0.67 l/min, P < 0.01). In contrast, no significant difference was found between measurements conducted with small lung volume breath‐holding (5.87 ± 0.53 l/min and 6.41 ± 0.75 l/min) and normal breathing. In conclusion, breath‐hold MR flow measurement using small lung volume by shallow inspiration can provide a blood flow quantification that is close to physiological blood flow. Magn Reson Med 45:346–348, 2001.