Evaluation of 1031 primary titanium nitride coated mobile bearing total knee arthroplasties in an orthopedic clinic
Authors:
- Stefan J M BreugemSBStefan J M Breugem
- Jeroen LinnartzJLJeroen Linnartz
- Inger SiereveltISInger Sierevelt
- Jon D BruijnJBJon D Bruijn
- Marcel J M DriessenMDMarcel J M Driessen
Open Access
- 18 December 2017
- journal article
- Published by Baishideng Publishing Group Inc. in World Journal of Orthopedics
- Vol. 8 (12) , 922-928
Abstract
To evaluate the influence of the titanium nitride (TiN) coating on the results of a total knee arthroplasty (TKA). A total of 910 patients (338 men; 572 woman), with a mean age of 65 (range 36-94) undergoing 1031 primary TKAs were assessed. Clinical evaluation and patient-reported outcomes were gathered one year after surgery. The questionnaires included the Knee injury and Osteoarthritis Outcome Score (KOOS)-Dutch version, Visual Analogue Scale (VAS) pain scores in rest and during active knee movement, VAS-satisfaction scores, and EQ-5D-3L health scores. This was aimed to assess the overall knee function and patient satisfaction, and to enable us to make a gross comparison to other TKAs. At a mean follow-up of 46 mo (range 1-92) the overall implant survival was 97.7% and 95.1% for any operative reason related to the implant. Twenty-three knees (2.2%) required revision surgery. Arthrofibrosis was the most common indication for a re-operation. The clinical evaluation and patient-reported outcomes revealed good to excellent patient satisfaction and function of the arthroplasty. The median postoperative VAS-pain scores on a scale of 0-100, at one year after surgery were 1 in rest and 2 during movement. The TiN coated, mobile bearing TKA results are excellent and similar to those of other widely used TKA designs. Residual pain of the knee remains a concern and the TiN coating in combination with the mobile bearing does not seem to be the simple solution to this problem. Future research will have to show that the coating gives a better survival than the cobalt chrome version.Keywords
This publication has 25 references indexed in Scilit:
- Mobile- versus fixed-bearing modern total knee replacements – which is the more patella-friendly design?Bone & Joint Research, 2013
- Revision Surgery for Patellofemoral Problems: Should We Always Resurface?Clinical Orthopaedics and Related Research, 2012
- The Epidemiology of Revision Total Knee Arthroplasty in the United StatesClinical Orthopaedics and Related Research, 2010
- LCS mobile-bearing total knee replacement. A 10-year's follow-up studyOrthopaedics & Traumatology: Surgery & Research, 2009
- The low contact stress mobile-bearing total knee replacementThe Journal of Bone and Joint Surgery. British volume, 2009
- Rotating Platform versus Fixed-bearing Total Knees: An In Vitro Study of WearClinical Orthopaedics and Related Research, 2008
- Neural Stimulation and Recording ElectrodesAnnual Review of Biomedical Engineering, 2008
- The Dutch version of the knee injury and osteoarthritis outcome score: A validation studyHealth and Quality of Life Outcomes, 2008
- Mobile-Bearing Knee Replacement: Clinical ResultsClinical Orthopaedics and Related Research, 2001
- Titanium nitride coatings in clinical dentistryJournal of Dentistry, 1992