In Vitro Performance of a Long-Range Surface Plasmon Hydrogel Aptasensor for Continuous and Real-Time Vancomycin Measurement in Human Serum
- 20 May 2024
- journal article
- research article
- Published by American Chemical Society (ACS) in ACS Applied Materials & Interfaces
- Vol. 16 (22) , 28162-28171
Abstract
This study investigated the suitability of surface modification for a long-range surface plasmon (LRSP) aptasensor using two different hydrogels, aiming at real-time monitoring of vancomycin (VCM) in undiluted serum and blood. Three different layer structures were formed on a gold surface of LRSP sensor chip using poly[2-methacryloyloxyethyl phosphorylcholine (MPC)-co-N-methacryloyl-(L)-tyrosinemethylester (MAT)] (PMM) and poly[MPC-co-2-ethylhexyl methacrylate (EHMA)-co-MAT] (PMEM). The peptide aptamer for VCM was immobilized in PMM and PMEM via MAT. Among four differently prepared sensor chips, the LRSP hydrogel aptasensor with PMM, referred to as the PMM hydrogel, exhibited the highest sensor output and superior antifouling properties. Following the optimization of the PMM hydrogel preparation conditions, the shelf life of the PMM hydrogel was determined to exceed 2 weeks, and the same sensor chip could be used for 102 days without significant performance deterioration. The PMM hydrogel was then applied for VCM measurement in undiluted serum in vitro, where it demonstrated a limit of detection of 0.098 μM and a dynamic range of 0.18-100 μM, covering the therapeutic range. Additionally, the PMM hydrogel enabled the continuous measurement of various VCM concentrations in serum without rinsing and showed a concentration-dependent output in undiluted blood. These findings underscore the potential of the PMM hydrogel for real-time and direct monitoring of VCM in body fluids.Keywords
Funding Information
- Japan Society for the Promotion of Science (JP18K13763, JP21H01335, JP23H03864)
- Japan Science and Technology Agency (JPMJPF2202)
- Cooperative Re-search Project of the Research Center for Biomedical Engineering
This publication has 37 references indexed in Scilit:
- Clinical and Pharmacokinetic Outcomes of Peak–Trough-Based Versus Trough-Based Vancomycin Therapeutic Drug Monitoring Approaches: A Pragmatic Randomized Controlled TrialEuropean Journal of Drug Metabolism and Pharmacokinetics, 2019
- Proteins and antibodies in serum, plasma, and whole blood—size characterization using asymmetrical flow field-flow fractionation (AF4)Analytical and Bioanalytical Chemistry, 2018
- The Whole Price of Vancomycin: Toxicities, Troughs, and TimeDrugs, 2017
- The Nephrotoxicity of VancomycinClinical Pharmacology & Therapeutics, 2017
- Innovative approaches to optimizing the delivery of vancomycin in individual patientsAdvanced Drug Delivery Reviews, 2014
- Vancomycin‐Associated Renal Dysfunction: Where Are We Now?Pharmacotherapy: The Journal of Human Pharmacology and Drug Therapy, 2014
- Long range surface plasmon and hydrogel optical waveguide field-enhanced fluorescence biosensor with 3D hydrogel binding matrix: On the role of diffusion mass transferBiosensors and Bioelectronics, 2010
- A pilot study to investigate patients reported knowledge, awareness, and beliefs on health care–associated infectionAmerican Journal of Infection Control, 2008
- Long Range Surface Plasmons for Observation of Biomolecular Binding Events at Metallic SurfacesPlasmonics, 2007
- Why do phospholipid polymers reduce protein adsorption?Journal of Biomedical Materials Research, 1998