Comparative Biomechanical Stability of Titanium Bone Fixation Systems in Metacarpal Fractures

Authors:
  • MMMichael McCarty
    Michael McCarty
  • TKThomas Katona
    Thomas Katona
  • KMKeith Moore
    Keith Moore
  • JJJ. Ryan Jackson
    J. Ryan Jackson
  • BEBarry L. Eppley
    Barry L. Eppley
  • RSRajiv Sood
    Rajiv Sood

Abstract
The biomechanical properties of apex bending and torsional rigidity of 11 different titanium mini- and microplates (Leibinger and Synthes) were examined to evaluate the effects of plate design and thickness and screw size and design. Two hundred sixteen fresh-frozen human cadaveric metacarpal bones underwent a midshaft transverse osteotomy followed by application of one of the following plates: Synthes 1.5-mm and 2.0-mm five-hole linear plates; Leibinger linear 1.2-mm 5-hole, 1.7-mm 4-hole, 2.3-mm 4-hole plates; or Leibinger three-dimensional 1.2-mm 4-hole, 1.2-mm 8-hole, 1.7-mm 4-hole, 1.7-mm 8-hole, 2.3-mm 4-hole, and 2.3-mm 8-hole plates. The specimens were subjected to a three-point bending test with apex dorsal or apex volar loading or torsional loading. Analysis of variance statistical analysis revealed that increasing plate thickness and, more significantly, three-dimensional design were associated with increased rigidity.