Oncological diagnosis using positron coincidence gamma camera with fluorodeoxyglucose in comparison with dedicated PET

Authors:
  • MTM Tian
    M Tian
  • NON Oriuchi
    N Oriuchi
  • THT Higuchi
    T Higuchi
  • STS Tanada
    S Tanada
  • KEK Endo
    K Endo

Abstract
The purpose of this study was to compare the utility of a dual-head positron coincidence detection gamma camera (PCD) with that of dedicated positron emission tomography (PET) in the imaging of various malignancies using 18F-fluorodeoxyglucose (FDG). 25 patients with known or suspected malignancies at various sites underwent imaging with both methods, and diagnostic performance on a lesion basis was compared. Tumour lesions were analyzed visually and semi-quantitatively using the ratio of tumour-to-background counts (T/B ratio). FDG PCD and FDG PET visually detected 34 (72.3%) lesions and 37 (78.7%) lesions, respectively. The mean T/B ratio and standard deviation (SD) of FDG PCD was 3.5±3.3, significantly lower than that of FDG PET (8.4±7.1, pp<0.01). Sensitivity between FDG PCD and FDG PET in lesions of more than 2.0 cm diameter showed no statistically significant difference. This study indicates that FDG imaging with a dual-head coincidence detection gamma camera can provide suitable diagnostic performance for lesions greater than 2.0 cm diameter, but performed significantly worse than dedicated PET for lesions smaller than this.