Surgical management of “Twiddler syndrome” in patients with deep brain stimulation: a technical note and review of the literature
Authors:
Open Access
- 25 February 2022
- journal article
- review article
- Published by Springer Nature in Acta Neurochirurgica
- Vol. 164 (4) , 1183-1186
Abstract
Background: The Twiddler syndrome (TS) describes a situation in which the implanted impulse generator (IPG) rotates several times around its own axis in the subcutaneous pocket. This can lead to severe mechanical damage of the leads and extensions and to dislocations. Method: Hereby, we report on a technique for revision surgery in patients diagnosed with Twiddler syndrome after undergoing previous deep brain stimulation (DBS) surgery. For revision surgery, the TYRX ™ Absorbable Antibacterial Envelope was used. Conclusion: The TS can be treated well with the envelope TYRX ™ Absorbable Antibacterial Envelope TYRX ™.Keywords
Funding Information
- Heinrich-Heine-Universität Düsseldorf
This publication has 10 references indexed in Scilit:
- Twiddler Syndrome without Lead Dislodgment Discovered by Remote MonitoringCase Reports in Cardiology, 2021
- Absorbable antibacterial envelope in the surgical management of Twiddler’s syndrome in a patient with gastric electric stimulator: a case reportDigestive Medicine Research, 2020
- The World-wide Randomized Antibiotic Envelope Infection Prevention (WRAP-IT) trial: Long-term follow-upHeart Rhythm, 2020
- The Unsuccessful Twiddler: A Case of Twiddler's Syndrome Without Deep Brain Stimulator Lead BreakageCureus, 2020
- Twiddler's syndrome: A rare hardware complication in spinal cord stimulationAsian Journal of Neurosurgery, 2018
- Use of an antimicrobial pouch/envelope in the treatment of Twiddler's syndromePacing and Clinical Electrophysiology, 2017
- Postoperative lead migration in deep brain stimulation surgery: Incidence, risk factors, and clinical impactPLOS ONE, 2017
- Twiddler’s syndrome in spinal cord stimulationActa Neurochirurgica, 2015
- Obese female patients have higher rates of lead dislodgement after ICD or CRT-D implantationInternational Journal of Cardiology, 2014
- Twiddler (or Not) Syndrome: Questioning etiology for an uncommon form of hardware malfunction in deep brain stimulationSurgical Neurology International, 2014