A simple EDXRF technique to analyse alloys
Authors:
- 15 January 2002
- journal article
- research article
- Published by EDP Sciences in The European Physical Journal Applied Physics
- Vol. 17 (1) , 81-84
Abstract
A simple EDXRF technique to obtain the relative concentrations of different elements present in a sample is described here. Only those elements have been considered whose characteristic X-rays fall within the sensitivity range of the X-ray detector that we used. A small computer program where the fundamental parameters such as photoionisation cross sections, fluorescence yields, Coster-Kronig transition rates, etc. have been used as inputs was written to calculate the relative concentrations of the elements. The technique used here requires only a single run with the sample and does not require any knowledge of the incoming X-ray flux or geometry of the experimental arrangement. Using this method, three alloys have been analysed in our existing EDXRF system.Keywords
This publication has 5 references indexed in Scilit:
- Reduction of scattered bremsstrahlung in a secondary target energy dispersive systemRadiation Physics and Chemistry, 2000
- Interpolated Dirac-Fock values of L-subshell x-ray emission rates including overlap and exchange effectsAtomic Data and Nuclear Data Tables, 1989
- XCOM :1987
- Atomic radiative and radiationless yields for K and L shellsJournal of Physical and Chemical Reference Data, 1979
- The Si(Li) X‐Ray energy analysis system: Operating principle and performanceX-Ray Spectrometry, 1972