Contactless Acquisition and Decoupling of Respiratory and Cardiac Signals using a 122 GHz FMCW Radar
Open Access
- 12 March 2026
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Journal of Biomedical and Health Informatics
- Vol. PP (21682194) , 1-14
Abstract
This paper presents a contactless method for the simultaneous monitoring of heart rate (HR) and respiratory rate (RR) using a Frequency Modulated Continuous Wave (FMCW) radar operating at 122.5 GHz. To address the challenges posed by spectral overlap and interference between cardiac and respiratory components, we introduce a signal separation and estimation framework based on Maximum Likelihood Estimation (MLE) combined with an orthogonal projection technique. The approach models the thoracic displacement captured by the radar as the superposition of periodic sources and applies spectral projection to isolate the fundamental frequencies and their harmonics. Theoretical performance is assessed through the derivation of the Cramér-Rao Lower Bound (CRLB) and extensive Monte Carlo simulations under varying SNR and sample size conditions. Experimental validation was conducted on 23 healthy volunteers, each providing two 20 second recordings, using a clinical-grade reference monitor. The method achieves high estimation accuracy, with a mean absolute error of 0.69 bpm for HR and 0.33 bpm for RR. The algorithm is capable of reconstructing the cardiac pulse waveform while preserving key morphological features such as the systolic peak and dicrotic notch. Furthermore, the use of the projection technique further enhanced estimation accuracy under nonphysiological conditions, including both elevated and reduced rates. These findings demonstrate the potential of the proposed method as a reliable tool for contactless monitoring of vital signs and support its future integration into biomedical applications.Keywords
This publication has 48 references indexed in Scilit:
- Reliability of an all-in-one wearable sensor for continuous vital signs monitoring in high-risk patients: the NIGHTINGALE clinical validation studyJournal of Clinical Monitoring and Computing, 2025
- Intelligent Heart Rate Extraction Method Based on Millimeter Wave RadarJournal of Shanghai Jiaotong University (Science), 2023
- Non-contact automatic vital signs monitoring of neonates in NICU using video camera imagingComputer Methods in Biomechanics and Biomedical Engineering: Imaging & Visualization, 2022
- Carbon nanotube materials for electrocardiographyRSC Advances, 2021
- Wavelet Analysis Based Noncontact Vital Signal Measurements Using mm-Wave Radar2020
- Vital-sign monitoring and spatial tracking of multiple people using a contactless radar-based sensorNature Electronics, 2019
- Novel signal processing method for vital sign monitoring using FMCW radarBiomedical Signal Processing and Control, 2017
- Non-contact video-based vital sign monitoring using ambient light and auto-regressive modelsPhysiological Measurement, 2014
- Multi-pitch estimationSignal Processing, 2008
- Three-dimensional reconstruction of cardiac displacement patterns on the chest wall during the P, QRS and T-segments of the ECG by laser speckle inteferometryMedical & Biological Engineering & Computing, 1989