Abstract
Background: Adrenal Insufficiency (AI), especially iatrogenic-AI, is a treatable cause of mortality. The difficulty in obtaining 9 am cortisol levels means samples are taken at suboptimal times, including a substantial proportion in the afternoon. Low afternoon cortisol levels often provoke short Synacthen tests (SSTs). It is important that this does not lead to patients misdiagnosed with AI, exposing them to the excess mortality and morbidity of inappropriate steroid replacement therapy. Methods: This retrospective study collected 60 178 cortisol results. Medical records, including subsequent SSTs of initial cortisol results measured after midday were reviewed. Results: Receiver operating characteristic analysis (area under the curve: 0.89) on 6531 suitable cortisol values showed that a limit of 234 nmol/L excludes AI on Abbott analyser platforms. In patients who have an afternoon cortisol <234 nmol/L, including both 30- and 60-min SST cortisol values prevents unnecessary glucocorticoid replacement therapy in 22.3% of individuals in this study. The Abbott analyser SST cortisol cutoffs used to define AI should be 366.5 nmol/L and 418.5 nmol/L at 30 and 60 min, respectively. All patients remained well subsequently with at least 1-year longitudinal follow-up.
Funding Information
  • National Institute for Health Research (DRF-2017-10-115)
  • Imperial Health Charity
  • Moulton Charitable Research Foundation
  • Higher Education Funding Council for England

This publication has 24 references indexed in Scilit: