Optimising Sample Preparation and Calibrations in EDXRF for Quantitative Soil Analysis
Open Access
Abstract
Energy-dispersive X-ray fluorescence spectrometry (EDXRF) is a rapid and inexpensive method for soil analysis; however, analytical results are influenced by particle size effects and spectral interferences. The objective of this study was to optimise sample preparation and calibrations to improve the accuracy of EDXRF for soil tests. Methods of sample preparation were compared by calculating the recoveries of 13 elements in four International Soil-Analytical Exchange (ISE) standards prepared as loose powder (LP), pressed pellet (PP), and pressed pellet with wax binder (PPB). A matching library (ML) was created and evaluated against the fundamental parameters (FP) calibration using 20 ISE standards. Additionally, EDXRF results of 41 tillage soils were compared with Inductively coupled plasma optical emission spectrometry (ICP-OES) results. The PPB had most recoveries within the acceptable range of 80–120%; conversely, PP yielded the poorest element recoveries. For the calibration, the ML provided better recoveries of Ni, Pb, Cu, Mg, S, P, and Cr; however, for Zn, and Mn, it had the opposite effect. Furthermore, EDXRF results compared with ICP-OES separated by soil texture class for Al, K, Mn, and Fe. In conclusion, the EDXRF is suitable for quantifying both trace elements and macronutrients in contaminated soils and has the potential to provide screening or prediction of soil texture in agriculture.Keywords
Funding Information
- Teagasc (6502)
- Queen's University Belfast (6502)
This publication has 47 references indexed in Scilit:
- Heavy metals, occurrence and toxicity for plants: a reviewEnvironmental Chemistry Letters, 2010
- X‐ray fluorescence analysis of Cr, As, Se, Cd, Hg, and Pb in soil using pressed powder pellet and loose powder methodsX-Ray Spectrometry, 2009
- A comparison of Cu, Pb, As, Cd, Zn, Fe, Ni and Mn determined by acid extraction/ICP–OES and ex situ field portable X-ray fluorescence analysesEnvironmental Pollution, 2006
- Energy-dispersive X-ray fluorescence systems as analytical tool for assessment of contaminated soilsJournal of Environmental Monitoring, 2004
- Determination of Cd, Cu, Pb and Zn in environmental samples: microwave-assisted total digestion versus aqua regia and nitric acid extractionAnalytica Chimica Acta, 2002
- Statistical Methods in Assessing AgreementJournal of the American Statistical Association, 2002
- Comparison of Techniques for the Analysis of Industrial Soils by Atomic SpectrometryInternational Journal of Environmental Analytical Chemistry, 1998
- The Bulk Analysis of Silicate Rocks by Portable X‐Ray Fluorescence: Effect of Sample Mineralogy in Relation to the Size of the Excited VolumeGeostandards Newsletter, 1997
- X-Ray fluorescence analysis utilizing the fundamental parameter method for the determination of the elemental composition in plant samplesAnalytica Chimica Acta, 1995
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986