Scaling Characteristics of the Magnetization Increments of Functionalized Nanoparticles Determined Using a Vibrating Sample Magnetometer for Liquid Magnetic Immunoassays
- 13 April 2018
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Sensors Journal
- Vol. 19 (20) , 8990-8994
Abstract
Here, scaling the normalized increment of dc magnetic susceptibility (Aχ/χ) to a characteristic logistic function provides a foundation for characterizing unknown quantities of biomarkers. We report the scaling behavior of magnetization increments in reagents composed of biofunctionalized magnetic nanoparticles (BMNs) conjugated with biotargets for liquid magnetic immunoassays. We measured hysteresis loops and obtained the increment of magnetization (AM) after binding the C-reactive protein (CRP) to Fe 3 O 4 -anti-CRP. The results show that AM increased when the concentration of CRP increased. There was enhanced AM generated from magnetic clusters composed of Fe 3 O 4 -anti-CRP-CRP. The present platform is robust, easy to use, and can be applied to assays for various biomolecules for tumor markers, viruses, and proteins.Keywords
Funding Information
- Ministry of Science and Technology, Taiwan (105-2112-M-003-012, 106-2112-M-003-002)
- National Taiwan Normal University (104J1A27, 105J1A27)
This publication has 27 references indexed in Scilit:
- Anti-CEA-functionalized superparamagnetic iron oxide nanoparticles for examining colorectal tumors in vivoDiscover Nano, 2013
- Magnetic Nanoparticles and microNMR for Diagnostic ApplicationsTheranostics, 2012
- Magnetic nanoparticles for high-sensitivity detection on nucleic acids via superconducting-quantum-interference-device-based immunomagnetic reduction assayJournal of Magnetism and Magnetic Materials, 2011
- Magnetic nanoparticle biosensorsWIREs Nanomedicine and Nanobiotechnology, 2010
- Gold-Coated Magnetic Particles for Solid-Phase Immunoassays: Enhancing Immobilized Antibody Binding Efficiency and Analytical PerformanceAnalytical Chemistry, 2005
- Magnetic Immunoassays Utilizing Magnetic Markers and a High-$T_rm c$SQUIDIEEE Transactions on Applied Superconductivity, 2005
- Application of high T/sub c/ SQUID magnetometer to biological immunoassaysIEEE Transactions on Applied Superconductivity, 2001
- Detection of Magnetic Nanoparticles with Superconducting Quantum Interference Device (SQUID) Magnetometer and Application to ImmunoassaysJapanese Journal of Applied Physics, 1999
- Determination of the binding reaction between avidin and biotin by relaxation measurements of magnetic nanoparticlesJournal of Magnetism and Magnetic Materials, 1999
- Assay of Plasma Insulin in Human Subjects by Immunological MethodsNature, 1959