Glucose and Lactate Miniaturized Biosensors for SECM-Based High-Spatial Resolution Analysis: A Comparative Study
- 24 August 2017
- journal article
- research article
- Published by American Chemical Society (ACS) in ACS Sensors
- Vol. 2 (9) , 1310-1318
Abstract
With the aim of developing miniaturized enzymatic biosensors suitable for in vitro diagnostic applications, such as monitoring of metabolites at single cell level, glucose and lactate biosensors were fabricated by immobilizing enzymes (glucose oxidase and lactate oxidase, respectively) on 10 μm Pt ultramicroelectrodes. These electrodes are meant to be employed as probes for scanning electrochemical microscopy (SECM), which is a unique technique for high-spatial-resolution electrochemical-based analysis. The use of enzymatic moieties improves sensitivity, time scale response, and information content of the microprobes; however, protein immobilization is a key step in the biosensor preparation that greatly affects the overall performance. A crucial aspect is the miniaturization of the sensing, preserving their sensitivity. In this work, we investigated the most common enzyme immobilization techniques. Several fabrication routes are reported and the main figures of merit, such as sensitivity, detection limit, response time, reproducibility, spatial resolution, biosensor efficiency, permeability, selectivity, and the ability to block electro-active interfering species, are investigated and compared. With the intent of using the microprobes for in vitro functional imaging of single living cells, we carefully evaluate the spatial resolution achieved by our modified electrodes on 2D SECM imaging. Metabolic activity of single MCF10A cells were obtained by monitoring the glucose concentrations in close proximity of single living cell, using the UME-based biosensor probes prepared. A voltage-switch approach was implemented to disentangle the topographical contribution of the cells enabling quantitative measurements of cellular uptakes.Keywords
Funding Information
- Associazione Italiana per la Ricerca sul Cancro
- Universit?? di Bologna
- Fondazione Italiana per la Ricerca sul Cancro
- Ministero della Salute (GR-2011-02348455)
This publication has 48 references indexed in Scilit:
- Amperometric detection of vesicular exocytosis from BON cells at carbon fiber microelectrodesElectrochimica Acta, 2014
- Topographical and electrochemical nanoscale imaging of living cells using voltage-switching mode scanning electrochemical microscopyProceedings of the National Academy of Sciences, 2012
- Review of analytical figures of merit of sensors and biosensors in clinical applicationsTrAC Trends in Analytical Chemistry, 2010
- Real-time monitoring of cell viability by its nanoscale height change with oxygen as endogenous indicatorChemical Communications, 2010
- Enzyme Immobilization Strategies and Electropolymerization Conditions to Control Sensitivity and Selectivity Parameters of a Polymer-Enzyme Composite Glucose BiosensorSensors, 2010
- A microelectrode biosensor for real time monitoring of l-glutamate releaseAnalytica Chimica Acta, 2009
- Nanoscale Imaging of Surface Topography and Reactivity with the Scanning Electrochemical MicroscopeAnalytical Chemistry, 2009
- Glucose and Lactate Biosensors for Scanning Electrochemical Microscopy Imaging of Single Live CellsAnalytical Chemistry, 2008
- Miniaturization of biosensors2002
- Electrochemical characterization of polypyrrole/glucose oxidase biosensor: Part I. Influence of enzyme concentration on the growth and properties of the filmBiosensors and Bioelectronics, 1996