Automated Measurement of Volume Flow in the Ascending Aorta Using MR Velocity Maps
Authors:
- R. André NiezenRNR. André Niezen
- Ernst E. van der WallEWErnst E. van der Wall
- Albert de RoosARAlbert de Roos
- Johan H. C. ReiberJRJohan H. C. Reiber
- 1 November 1998
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Computer Assisted Tomography
- Vol. 22 (6) , 904-911
Abstract
An automated contour detection algorithm was developed for the objective and reproducible quantitative analysis of velocity-encoded MR studies of the ascending aorta. The only user interaction required is the manual definition of a center point inside the cross-section of the aorta in one of the available images. The automated contour detection algorithm detects an initial model contour in this image and subsequently corrects for motion and deformation of the aortic cross-section in each of the acquired images over the complete cardiac cycle using dynamic programming techniques. Integrating the flow velocity values for each pixel within the detected contour results in an instantaneous flow value. Next, by integrating the instantaneous flow values for each acquired phase over the complete cardiac cycle, left ventricular stroke volume measurement could be obtained. The results of the automated method were compared with results derived from manually traced contours in MR studies from 11 healthy volunteers. An excellent agreement in stroke volume measurements was observed: signed difference 0.61 ± 1.15%. Inter- and intraobserver variabilities were <2% for both manual and automated image analysis methods. Manual tracing of contours required on the order of 10 min; the analysis time for automated contour detection was <6 s/study. The present contour detection allows fast and reliable left ventricular stroke volume measurements from aortic flow studies using velocity-encoded MR studies in healthy volunteers. Further study is required to assess the accuracy and reproducibility of the algorithm in patients with aortic valve disease.Keywords
This publication has 11 references indexed in Scilit:
- Comparison Between Manual and Semiautomated Analysis of Left Ventricular Volume Parameters from Short-Axis MR ImagesJournal of Computer Assisted Tomography, 1997
- Cardiovascular applications of magnetic resonance flow and velocity measurementsJournal of Magnetic Resonance Imaging, 1996
- Exercise‐related changes in aortic flow measured with spiral echo‐planar MR velocity mappingJournal of Magnetic Resonance Imaging, 1995
- Magnetic resonance measurement of velocity and flow: Technique, validation, and cardiovascular applicationsAmerican Heart Journal, 1993
- Quantification of aortic regurgitation by magnetic resonance velocity mappingAmerican Heart Journal, 1993
- Severity of aortic regurgitation: interstudy reproducibility of measurements with velocity-encoded cine MR imaging.Radiology, 1992
- Right and left ventricular stroke volume measurements with velocity-encoded cine MR imaging: in vitro and in vivo validation.American Journal of Roentgenology, 1991
- Vena caval flow: assessment with cine MR velocity mapping.Radiology, 1990
- Quantitation of antegrade and retrograde blood flow in the human aorta by magnetic resonance velocity mappingAmerican Heart Journal, 1989