Source apportionment and quantification of liquid and headspace leaks from closed system drug-transfer devices via Selected Ion Flow Tube Mass Spectrometry (SIFT-MS)
Authors:
Open Access
- 4 November 2021
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 16 (11) , e0258425
Abstract
A system to differentiate and quantify liquid and headspace vapor leaks from closed system drug-transfer devices (CSTDs) is presented. CSTDs are designed to reduce or eliminate hazardous drug (HD) exposure risk when compounding and administering HDs. CSTDs may leak liquid, headspace, or a mixture of the two. The amount of HD contained in liquid and headspace leaks may be substantially different. Use of a test solution containing two VOCs with differences in ratios of VOC concentrations in the headspace and liquid enables source apportionment of leaked material. SIFT-MS was used to detect VOCs from liquid and headspace leaks in the vapor phase. Included in this report is a novel method to determine the origin and magnitude of leaks from CSTDs. A limit of leak detection of 24 μL of headspace vapor and 0.14 μL of test liquid were found using Selected Ion Flow Tube Mass Spectrometry (SIFT-MS).Keywords
This publication has 19 references indexed in Scilit:
- Evaluation of a closed-system cytotoxic transfer device in a pharmaceutical isolatorJournal of Oncology Pharmacy Practice, 2014
- Leakage from closed-system transfer devices as detected by a radioactive tracerAmerican Journal of Health-System Pharmacy, 2013
- Multiple-test assessment of devices to protect healthcare workers when administering cytotoxic drugs to patientsJournal of Oncology Pharmacy Practice, 2011
- NIOSH Definition of Closed-System Drug-Transfer DevicesAnnals of Occupational Hygiene, 2009
- Contamination Comparison of Transfer Devices Intended for Handling Hazardous DrugsHospital Pharmacy, 2008
- Spill and Leakage Using a Drug Preparation System Based on Double-Filter TechnologyAnnals of Occupational Hygiene, 2008
- Validation protocol and analytical quality in biological monitoring of occupational exposure to antineoplastic drugsToxicology Letters, 2006
- Uncertainty in source partitioning using stable isotopesOecologia, 2001
- SIFT studies of the reactions of H3O+, NO+ and O2+ with several ethersInternational Journal of Mass Spectrometry and Ion Processes, 1998
- Reactions of H3O+ and OH− ions with some organic molecules; applications to trace gas analysis in airInternational Journal of Mass Spectrometry and Ion Processes, 1995