Abstract
Rationale: Measurements of oxygen‐17 excess ( 17 O excess ) in carbonates have become of great importance. However, to compare results obtained by different laboratories, it is necessary to normalize them to international standards. With this purpose in mind, we measured accurate and high precision δ 17 O and 17 O excess values for NBS19 and NBS18, two international standards, for which δ 18 O values are already widely used as the references for carbonates. Methods: The measurements are based on isotopic exchange at steady state between O 2 and CO 2 over hot platinum sponge at 750°C. Dual‐inlet isotope ratio mass spectrometry (IRMS) measurements of the δ 17 O and δ 18 O values of this O 2 allow δ 17 O values of CO 2 to be obtained with very high precision (0.01 to 0.03 ‰) and, correspondingly, accurate 17 O excess values with a precision of 5 per meg. Results: We measured, for the first time, the δ 17 O values and 17 O excess of CO 2 liberated from NBS19 at 25°C (39.196 ± 0.026; 20.276 ± 0.015 ‰ and −227 ± 4 per meg, respectively) and NBS18 (17.591 ± 0.041 ‰; 9.253 ± 0.018 ‰ and 3 ± 5 per meg, respectively). The values are given versus VSMOW. Conclusions: Accurate values of δ 17 O and 17 O excess of the international standards NBS19‐CO 2 and NBS18‐CO 2 are now available. The new values should be used for normalization of measured oxygen isotope ratios of carbonates to allow meaningful comparison of results among different laboratories. Copyright © 2015 John Wiley & Sons, Ltd.

This publication has 16 references indexed in Scilit: