An Electrochemical Detector for Liquid Chromatography with Picogram Sensitivity
Authors:
- 1 May 1973
- journal article
- research article
- Published by Taylor & Francis in Analytical Letters
- Vol. 6 (5) , 465-477
Abstract
A simple liquid chromatograph with an extremely sensitive electrochemical detector is described. The detector, which can be constructed for lees than $5.00 (excluding electronics), has an actual dead volume of < 1 μl and operates well in the 50–100 picogram region. Routine quantitative work at the 5–10 nanogram level is described. Potential applications to the analysis of biogenic amines are discussed.Keywords