Randomised comparison of manual compression and FemoSealª vascular closure device for closure after femoral artery access coronary angiography: the CLOSure dEvices Used in everyday Practice (CLOSE-UP) study
Authors:
- Niels R. HolmNHNiels R. Holm
- Birthe SindbergBSBirthe Sindberg
- Mia SchouMSMia Schou
- Michael MaengMMMichael Maeng
- Anne KaltoftAKAnne Kaltoft
- Morten BøttcherMBMorten Bøttcher
- Lars R. KrusellLKLars R. Krusell
- Jakob HjortJHJakob Hjort
- Leif ThuesenLTLeif Thuesen
- Christian J. TerkelsenCTChristian J. Terkelsen
- Evald H. ChristiansenECEvald H. Christiansen
- Hans Erik BøtkerHBHans Erik Bøtker
- 1 June 2014
- journal article
- research article
- Published by Europa Digital & Publishing in EuroIntervention
- Vol. 10 (2) , 183-190
Abstract
To compare in a randomised trial the safety and efficacy of the FemoSeal vascular closure device (VCD) versus manual compression (MC) after femoral access coronary angiography (CAG). In 13 months, 1,014 patients were included and 1,001 patients entered analysis. Median [interquartile range] closure time was 8.0 [6-10] minutes after MC versus 1.0 [1-1] minute (p<0.0001) for the FemoSeal VCD. Bed rest for one hour after the closure procedure was recommended in both groups. The primary endpoint of incidence of large groin haematoma was 6.7% in the MC group vs. 2.2% (p=0.002) in the FemoSeal group. The combined endpoint of 14-day adverse vascular events occurred in 1.0% in the MC group vs. 0.6% in the FemoSeal VCD group (p=0.7). Manual compression (OR 3.3, 95% CI: 1.5-7.2, p=0.002), female gender (OR 2.1, 95% CI: 1.1-3.9, p=0.018), and multiple punctures (OR 10.5, 95% CI: 3.2-34.3, p=0.001) were identified as independent predictors of adverse events and large haematomas. Closure of femoral access after coronary angiography by the FemoSeal vascular closure device was safe, faster, and associated with significantly fewer in-hospital large haematomas as compared to closure by manual compression.Keywords
This publication has 21 references indexed in Scilit:
- Pain and discomfort in closure of femoral access coronary angiography. The CLOSuredEvices Used in everyday Practice (CLOSE-UP) pain sub studyEuropean Journal of Cardiovascular Nursing, 2013
- Safety and efficacy of femoral artery closure with the FemoSeal® device after coronary angiography using a 7 French sheathPerfusion, 2011
- Prevalence of bruising at the vascular access site one week after elective cardiac catheterisation or percutaneous coronary interventionJournal of Clinical Nursing, 2011
- Impact of Femoral Vascular Closure Devices and Antithrombotic Therapy on Access Site Bleeding in Acute Coronary SyndromesCirculation: Cardiovascular Interventions, 2010
- Vascular Closure Device Failure: Frequency and ImplicationsCirculation: Cardiovascular Interventions, 2009
- A retrospective study of the management of retro-peritoneal bleeding after percutaneous coronary intervention in the era of combination anti-platelet therapyJournal of Cardiovascular Medicine, 2008
- Randomized Trial Comparing Same-Day Discharge With Overnight Hospital Stay After Percutaneous Coronary InterventionCirculation, 2007
- Hours During and After Coronary Intervention and AngiographyClinical Nursing Research, 2006
- Vascular Complications with Newer Generations of Angioseal Vascular Closure DevicesJournal of Interventional Cardiology, 2006
- Assessment of the safety and efficacy of the DUETT vascular hemostasis device: Final results of the safe and effective vascular hemostasis (SEAL) trialAmerican Heart Journal, 2002