In vivo porcine liver radiofrequency ablation with simultaneous MR temperature imaging
Authors:
- Jerry JarrardJJJerry Jarrard
- Kim Butts PaulyKPKim Butts Pauly
Open Access
- 28 February 2006
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 23 (4) , 578-584
Abstract
Purpose: To demonstrate in vivo MR‐guided temperature mapping during radiofrequency (RF) ablation of the liver with a commercially available RF generator modified to allow simultaneous RF treatment and MRI.Materials And Methods: A commercial RF generator was modified using passive filtering to allow the continuous application of the treatment current during MRI studies. A total of six ablations were performed with the device in vivo in three porcine livers, and imaging was concurrently performed using one of two different temperature mapping strategies.Results: MR images acquired during RF ablation demonstrated no noticeable interference from the RF ablation device, which was operated at clinically relevant power levels. Temperature maps showed areas of heating that were consistent with the dimensions of the RF ablation probe, with some asymmetry (likely depending on the orientation of the probe and heat propagation effects), and some differences in heating‐spot area stability depending on the specific temperature mapping strategy used. Lesions were visualized on post‐ablation imaging and sectioning.Conclusion: The feasibility of performing RF ablation with a modified commercial RF generator simultaneously with MRI was demonstrated. Interference‐free MR temperature maps were produced with both variable respiratory motion and mechanical ventilation, and showed the extent of heating as the ablation progressed. J. Magn. Reson. Imaging 2006.Keywords
This publication has 14 references indexed in Scilit:
- MR-guided radiofrequency ablation of hepatic malignancies at 1.5 T: Initial resultsJournal of Magnetic Resonance Imaging, 2004
- Triggered, navigated, multi‐baseline method for proton resonance frequency temperature mapping with respiratory motionMagnetic Resonance in Medicine, 2003
- Radio-frequency-induced thermal lesions: Subacute magnetic resonance appearance and histological correlationJournal of Magnetic Resonance Imaging, 2003
- Reduction of Electronic Noise From Radiofrequency Generator During Radiofrequency Ablation in Interventional MRIJournal of Computer Assisted Tomography, 2002
- Percutaneous Radiofrequency Ablation of Liver Tumors with the LeVeen Probe: Is Roll-off Predictive of Response?Journal of Vascular and Interventional Radiology, 2001
- Temperature quantification using the proton frequency shift technique: In vitro and in vivo validation in an open 0.5 tesla interventional MR scanner during RF ablationJournal of Magnetic Resonance Imaging, 2001
- Radio-frequency-induced thermoablation: monitoring with T1-weighted and proton-frequency-shift MR imaging in an interventional 0.5-T environment.Radiology, 1998
- The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kindsMedical Physics, 1996
- Noninvasive MRI Thermometry with the Proton Resonance Frequency (PRF) Method: In Vivo Results in Human MuscleMagnetic Resonance in Medicine, 1995
- Non-invasive thermometry using magnetic resonance diffusion imaging: Potential for application in hyperthermic oncologyInternational Journal of Hyperthermia, 1992