Method of thermoluminescent measurement of radiation doses from micrograys up to a megagray with a single LiF:Mg,Cu,P detector
Open Access
- 3 November 2010
- journal article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 144 (1-4) , 543-547
Abstract
On the basis of the newly discovered behaviour of LiF:Mg,Cu,P detectors at high and ultra-high doses, a new method of thermoluminescence (TL) measurement of radiation doses ranging from micrograys up to a megagray, has been recently developed at the Institute of Nuclear Physics (IFJ). The method is based on the relationship between the TL signal, integrated in the given temperature range and dose. It is quantified by a parameter called the ‘ultra-high temperature ratio’. It has been demonstrated that this new method can measure radiation doses in the range of about 1 µGy to 1 MGy, using a single LiF:Mg,Cu,P detector. This method was recently successfully blindly tested for 10 MeV electrons up to doses of 200 kGy. It can be used for dosimetry in high-energy accelerators, especially in the Large Hadron Collider at CERN, and has great potential for accident dosimetry in particular.Keywords
This publication has 13 references indexed in Scilit:
- The response of TL lithium fluoride detectors to 24 GeV/c protons for doses ranging up to 1 MGyRadiation Measurements, 2010
- Behaviour of LiF:Mg,Cu,P and LiF:Mg,Ti thermoluminescent detectors for electron doses up to 1MGyRadiation Measurements, 2010
- Development of a Method for Passive Measurement of Radiation Doses at Ultra-High Dose RangeIEEE Transactions on Nuclear Science, 2009
- The response of different types of TL lithium fluoride detectors to high-energy mixed radiation fieldsRadiation Measurements, 2008
- Characteristics of LiF:Mg,Cu,P thermoluminescence at ultra-high dose rangeRadiation Measurements, 2008
- Workplace characterisation in mixed neutron-gamma fields, specific requirements and available methods at high-energy acceleratorsRadiation Protection Dosimetry, 2007
- The problems associated with the monitoring of complex workplace radiation fields at European high-energy accelerators and thermonuclear fusion facilitiesRadiation Protection Dosimetry, 2007
- High-dose characterization of different LiF phosphorsRadiation Measurements, 2007
- Theory of thermoluminescenceRadiation Measurements, 2006
- LiF:Mg,Ti Versus LiF:Mg,Cu,P: The Competition Heats UpRadiation Protection Dosimetry, 1993