Overlapping Ablation Using a Coaxial Radiofrequency Electrode and Multiple Cannulae System: Experimental Study in ex-Vivo Bovine Liver
Open Access
- 1 January 2003
- journal article
- Published by The Korean Society of Radiology in Korean Journal of Radiology
- Vol. 4 (2) , 117-23
Abstract
To assess the sizes and configurations of thermal zones after overlapping ablations using a coaxial radiofrequency (RF) electrode and multiple cannulae in ex-vivo bovine liver. For ablation procedures, a coaxial RF electrode and introducer set was used. Employing real-time ultrasound guidance and overlapping techniques in explanted, fresh bovine liver, we created five kinds of thermal zones with one (n=10), two (n=8), four (n=3), and six ablation spheres (n=3). Following ablation, MR images were obtained and the dimensions of all thermal zones were measured on the longitudinal or transverse section of specimens. The shape of the composite ablation zones was evaluated using three-dimensional MR image reconstruction. At gross pathologic examination of ten single-ablation zones (spheres), the long-axis (transverse) and short-axis lengths of zones ranged from 3.7 to 4.4 (mean, 4.1) cm and from 3.5 to 4.0 (mean, 3.7) cm, respectively. The long-axis (transverse) and short-axis lengths of double-ablation zones (cylinders) with 23% overlap ranged from 7.0 to 7.7 (mean, 7.3) cm and from 3.0 to 3.9 (mean, 3.5) cm, respectively; those with 58% overlap ranged from 6.0 to 6.4 (mean, 6.2) cm and from 3.8 to 4.6 (mean, 4.3) cm, respectively. The long-axis (diagonal) and short-axis lengths on a transverse section of four-ablation zones (cakes) ranged from 8.5 to 9.7 (mean, 9.1) cm and from 3.0 to 4.1 (mean, 3.7) cm, respectively. Gross pathologic examination of three composite six-ablation zones (spheres) showed that the long-axis (diagonal) and short-axis lengths of zones ranged from 9.0 to 9.9 (mean, 9.4) cm and from 6.8 to 7.5 (mean, 7.2) cm, respectively. T2-weighted MR images depicted low-signal thermal zones containing multiple curvilinear and spotty regions of hyperintensity. Using a coaxial RF electrode and multiple cannulae, together with ultrasound guidance and precise overlapping ablation techniques, we successfully created predictable thermal zones in ex-vivo bovine liver.Keywords
This publication has 16 references indexed in Scilit:
- Radiofrequency Thermal AblationAmerican Journal of Roentgenology, 2001
- MR Imaging-Histopathologic Correlation of Radiofrequency Thermal Ablation Lesion in a Rabbit Liver Model: Observation during Acute and Chronic StagesKorean Journal of Radiology, 2001
- Radiofrequency Liver AblationAmerican Journal of Roentgenology, 2001
- Radiofrequency Ablation of the LiverAmerican Journal of Roentgenology, 2001
- Creation of Radiofrequency Lesions in a Porcine ModelAmerican Journal of Roentgenology, 2000
- Thermal Ablation Therapy for Focal MalignancyAmerican Journal of Roentgenology, 2000
- Treatment of VX2 liver tumor in rabbits with "wet" electrode mediated radio-frequency ablationEuropean Radiology, 2000
- Radiofrequency Thermal Ablation of Hepatocellular CarcinomasKorean Journal of Radiology, 2000
- Ablation of liver tumors using percutaneous RF therapy.American Journal of Roentgenology, 1998
- Volume determinations using computed tomographyAmerican Journal of Roentgenology, 1982