Abstract
Online radiation dose monitoring is critical for the safe operation of accelerators. Although commercial dose monitors are well-developed, integrating an ionization chamber directly within a small-sized Betatron magnet remains challenging. In this study, we designed an air ionization chamber tailored for real-time dose monitoring in a small-sized Betatron. We selected aluminum for the chamber wall based on structural and integration requirements, designed the cavity geometry, and developed the associated charge collection and sampling circuits. Using a standard reference PTW ionization chamber, we calibrated the output voltage of the chamber against X-ray dose rates and conducted stability tests. The results show that there is a very good linear relationship between the output voltage of the ionization chamber and the X-ray dose rate. The relative standard deviation of the dose rate data within a 10 min working cycle is 3.25%, and the dose rate data shows good consistency with the standard reference ionization chamber. The ionization chamber can ensure operational safety for a small-sized Betatron and offer guidance for similar applications.
Funding Information
  • National Natural Science Foundation of China (42374226)
  • Key Research and Development Project of Jiangxi Province (20232BBE50013)
  • Natural Science Foundation of Jiangxi Province (20243BCE51132, 20252BAC200169)
  • National Key Laboratory of Uranium Resource Exploration-Mining and Nuclear Remote Sensing (2024HDX10)

This publication has 19 references indexed in Scilit: