Physiological Flow Simulation in Residual Human Stenoses After Coronary Angioplasty
- 20 March 2000
- journal article
- research article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 122 (4) , 310-320
Abstract
To evaluate the local hemodynamic implications of coronary artery balloon angioplasty, computational fluid dynamics (CFD) was applied in a group of patients previously reported by [Wilson et al. (1988), 77, pp. 873–885] with representative stenosis geometry post-angioplasty and with measured values of coronary flow reserve returning to a normal range (3.6±0.3). During undisturbed flow in the absence of diagnostic catheter sensors within the lesions, the computed mean pressure drop was only about 1 mmHg at basal flow, and increased moderately to about 8 mmHg for hyperemic flow. Corresponding elevated levels of mean wall shear stress in the midthroat region of the residual stenoses, which are common after angioplasty procedures, increased from about 60 to 290 dynes/cm2 during hyperemia. The computations indicated that the pulsatile flow field was principally quasi-steady during the cardiac cycle, but there was phase lag in the pressure drop−mean velocity relation. Time-averaged pressure drop values, were about 20 percent higher than calculated pressure drop values, for steady flow, similar to previous in vitro measurements by Cho et al. (1983). In the throat region, viscous effects were confined to the near-wall region, and entrance effects were evident during the cardiac cycle. Proximal to the lesion, velocity profiles deviated from parabolic shape at lower velocities during the cardiac cycle. The flow field was very complex in the oscillatory separated flow reattachment region in the distal vessel where pressure recovery occurred. These results may also serve as a useful reference against catheter-measured pressure drops and velocity ratios (hemodynamic endpoints) and arteriographic (anatomic) endpoints post-angioplasty. Some comparisons to previous studies of flow through stenoses models are also shown for perspective purposes. [S0148-0731(00)00304-6]
Keywords
This publication has 32 references indexed in Scilit:
- A Scaling Law for Wall Shear Rate Through an Arterial StenosisJournal of Biomechanical Engineering, 1994
- Flow Measurements in a Highly Curved Atherosclerotic Coronary Artery Cast of ManJournal of Biomechanical Engineering, 1992
- Experimental Investigation of Branch Flow Ratio, Angle, and Reynolds Number Effects on the Pressure and Flow Fields in Arterial Branch ModelsJournal of Biomechanical Engineering, 1985
- Experimental study of pulsatile and steady flow through a smooth tube and an atherosclerotic coronary artery casting of manJournal of Biomechanics, 1983
- Initiation of turbulence in models of arterial stenosesJournal of Biomechanics, 1979
- The pulsed Doppler coronary artery catheter preliminary report of a new technique for measuring rapid changes in coronary artery flow velocity in man.Circulation, 1977
- Quantitative coronary arteriography: estimation of dimensions, hemodynamic resistance, and atheroma mass of coronary artery lesions using the arteriogram and digital computation.Circulation, 1977
- Analysis of pulsatile, viscous blood flow through diseased coronary arteries of manJournal of Biomechanics, 1977
- Steady laminar flow through modelled vascular stenosesJournal of Biomechanics, 1976
- Shear-Layer Flow Regimes and Wave Instabilities and Reattachment Lengths Downstream of an Abrupt Circular Channel ExpansionJournal of Applied Mechanics, 1972