Development of a fully automatic separation system coupled with online ICP-MS for measuring rare earth elements in seawater
Open Access
- 22 August 2022
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in RSC Advances
- Vol. 12 (37) , 24003-24013
Abstract
Rare earth elements (REEs) are useful geological indicators of marine geochemistry. However, extremely low concentrations (sub-ng L−1) and high-salt matrices result in inefficient measurements. A fully automatic separation system (ELSPE-2 Precon) is used in the online determination of ultra-trace REEs in seawater using inductively coupled plasma mass spectrometry. This system mainly comprises three sections: (i) an auto-sampler (eas-2A) with 120 positions; (ii) a poly(styrene-divinylbenzene) resin column (Prin-Cen Col007) with iminodiacetic and ethylenediaminetriacetic acid functional groups to eliminate the high-salt matrix (e.g., Na, Ca, K, Mg, Al, Ba, Fe, Sr, P, and S) and preserve the target REEs; and (iii) a Trp002 cleanup column for the reduction of the reagent and procedural blank values. The detection limits (3σ) were in the range 0.002 (Dy)–0.097 ng L−1 (La), and the long-term reproducibility (8 h) was between 80% and 120% for all REEs in a 3.5% NaCl matrix solution. The accuracy of this method was verified using a seawater reference material (NASS-6), and the measured REE concentrations were consistent with those previously reported. The proposed online system was used to investigate coastal water samples with varying salinities from the Pearl River Estuary (Guangdong, China). Variations in the REE distribution patterns of different layers of seawater were observed, which could be due to the mixing of potentially light rare earth element-enriched bottom seawater. Moreover, a positive Gd anomaly in river water and seawater might be attributed to anthropogenic pollution from hospitals and the pharmaceutical industry.Keywords
Funding Information
- National Natural Science Foundation of China (41873072, 41521001, 4202010400)
- Natural Science Foundation of Hubei Province (2020CFA045)
This publication has 62 references indexed in Scilit:
- Online preconcentration ICP‐MS analysis of rare earth elements in seawaterGeochemistry, Geophysics, Geosystems, 2012
- Multi‐Element Determination of Trace Elements in Natural Water Reference Materials by ICP‐SFMS after Tm Addition and Iron Co‐precipitationGeostandards and Geoanalytical Research, 2010
- Determination of REEs in seawater by ICP-MS after on-line preconcentration using a syringe-driven chelating columnTalanta, 2009
- Retro-review of flow-injection analysisTrAC Trends in Analytical Chemistry, 2008
- Determination of trace rare earth elements in environmental samples by low temperature electrothermal vaporization inductively coupled plasma mass spectrometry after synergistic extraction with dimethylheptyl methyl phosphate and 1-phenyl-3-methyl-4-benzoyl-pyrazalone-5Analytica Chimica Acta, 2007
- Multielement determination of trace metals in seawater by ICP-MS with aid of down-sized chelating resin-packed minicolumn for preconcentrationTalanta, 2007
- Dissolved rare earth elements in river waters draining karst terrains in Guizhou Province, ChinaAquatic Geochemistry, 2006
- Determination of lanthanides in rock samples by inductively coupled plasma mass spectrometry using thorium as oxide and hydroxide correction standardSpectrochimica Acta Part B: Atomic Spectroscopy, 2003
- The rainbow vent fluids (36°14′N, MAR): the influence of ultramafic rocks and phase separation on trace metal content in Mid-Atlantic Ridge hydrothermal fluidsChemical Geology, 2002
- The rare earth elements in seawaterNature, 1982