Abstract
The relation between the voltage induced in an open sensor and the magnetic moment of a sample during pulsed field magnetization measurements has been calculated. The calculation takes into account the size of a sample, its internal flux distribution, and its position with respect to the center of the sensor. It has been found that the behavior of the response signal with respect to the sample position is the same for homogeneously magnetized Ni samples and high temperature superconductors. Because of this universality, it is possible to calibrate the open sensor even in the case of an unfavorable sample geometry. The calculations have been experimentally verified by using Ni samples with different geometries. Our results can easily be extended to samples with an arbitrary local field distribution.Keywords
This publication has 9 references indexed in Scilit:
- Experimental techniques for pulsed magnetic fieldsPhysica B: Condensed Matter, 1996
- A probe for magnetization measurements of thin superconducting films in pulsed high magnetic fieldsReview of Scientific Instruments, 1995
- Anisotropic critical currents in Ba2YCu3O7 analyzed using an extended Bean modelApplied Physics Letters, 1989
- Critical-state model for the determination of critical currents in disk-shaped superconductorsJournal of Applied Physics, 1979
- Magnetization of High-Field SuperconductorsReviews of Modern Physics, 1964
- Magnetization and Critical SupercurrentsPhysical Review B, 1963
- Theory of Flux Creep in Hard SuperconductorsPhysical Review Letters, 1962
- Critical Persistent Currents in Hard SuperconductorsPhysical Review Letters, 1962
- Magnetization of Hard SuperconductorsPhysical Review Letters, 1962