The Effect of Scatter and Attenuation on Aerosol Deposition as Determined by Gamma Scintigraphy
Authors:
- Zhenghong LeeZLZhenghong Lee
- Marc S. BerridgeMBMarc S. Berridge
- A. Dennis NelsonANA. Dennis Nelson
- Donald L. HealdDHDonald L. Heald
- 1 June 2001
- journal article
- research article
- Published by SAGE Publications in Journal of Aerosol Medicine
- Vol. 14 (2) , 167-183
Abstract
Gamma scintigraphy is often used to quantify deposition patterns from aerosol inhalers. The errors caused by scatter and tissue attenuation in planar Tc-99m gamma scintigraphy were investigated based on the data collected from four subjects in this study. Several error correction methods were tested. The results from two scatter correction methods, Jaszczak's method and factor analysis of dynamic sequences (FADS), were similar. Scatter accounted for 20% of raw data in the whole lung, 20% in the oropharynx, and 43% in the central airways and esophagus. Three attenuation correction methods were investigated and compared. These were: uniform attenuation correction (UAC), a known method used for inhalation drug imaging work; the broad-beam attenuation correction used for organ imaging in nuclear medicine; and a narrow-beam inhomogeneous tissue attenuation correction proposed in this study. The three methods differed significantly (p < 0.05), but all indicated that attenuation is a severe quantification problem. The narrow beam attenuation correction with scatter correction, showed that raw data underestimated tracer deposition by 44% in the lung, 137% in the oropharynx, and 153% in the trachea/esophageal region. To quantify aerosol lung deposition using planar scintigraphy even in relative terms, corrections are necessary. Much of the literature concerning quantified aerosol dose distributions measured by gamma scintigraphy needs to be interpreted carefully.Keywords
This publication has 27 references indexed in Scilit:
- A Technique for Simulating Radionuclide Images from the Aerosol Deposition Pattern in the Airway TreeJournal of Aerosol Medicine, 1997
- Tissue Attenuation Corrections in Gamma ScintigraphyJournal of Aerosol Medicine, 1997
- Scintigraphic Assessment of the Oropharyngeal and Nasal Depositions of Fusafungine from a Pressurized Inhaler and from a Novel Pump Spray DeviceJournal of Pharmacy and Pharmacology, 1995
- Description of Pulmonary Deposition of Radiolabeled Aerosol by Airway Generation Using a Conceptual Three Dimensional Model of Lung MorphologyJournal of Aerosol Medicine, 1995
- Lung deposition of budesonide inhaled via Turbuhaler: a comparison with terbutaline sulphate in normal subjectsEuropean Respiratory Journal, 1994
- Target apex-seeking in factor analysis of medical image sequencesPhysics in Medicine & Biology, 1993
- Pulmonary Deposition of Inhaled Terbutaline: Comparison of Scanning Gamma Camera and Urinary Excretion MethodsJournal of Pharmaceutical Sciences, 1992
- Scatter correction in planar imaging and SPECT by constrained factor analysis of dynamic structures (FADS)Physics in Medicine & Biology, 1990
- A technique for using CT images in attenuation correction and quantification in SPECTNuclear Medicine Communications, 1989
- A technique for the absolute measurement of activity using a gamma camera and computerPhysics in Medicine & Biology, 1979