Three-Dimensional Fluoroscopy For Evaluation of Articular Reduction and Screw Placement in Calcaneal Fractures
- 1 November 2007
- journal article
- research article
- Published by SAGE Publications in Foot & Ankle International
- Vol. 28 (11) , 1165-1171
Abstract
Background: Anatomic reconstruction of the posterior calcaneal facet after intra-articular fracture is one of the critical factors in achieving a good functional result. Intraoperative evaluation of fracture reduction and implant placement relies on direct view by standard fluoroscopy. We hypothesized that three-dimensional (3D) fluoroscopy is more accurate than conventional fluoroscopy, and equivalent to CT for determining fracture reduction and screw position in calcaneal fractures. Methods: A Sanders type IIB fracture pattern was created in eight embalmed lower extremity cadaver specimens. First, the posterior facet was reduced with a step-off of 0 mm to 2 mm in 0.5-mm increments. All specimens had two dimensional (2D) fluoroscopy, 3D fluoroscopy with an Iso-C3D, and a CT scan. Next, screws were placed so they protruded into the subtalar and calcaneocuboid joint and through the medial wall. All specimens were imaged again. Three observers evaluated all imaging studies, and the sensitivities and specificities of each modality were determined. Results: Both the Iso-C3D and the CT were more specific for anatomic reduction (75% and 100%, respectively) than fluoroscopy (62%). For the malreduced trials, the Iso-C3D and CT were both 100% sensitive, and the sensitivity of fluoroscopy was 63% ( p < 0.001 for both). For the intra-articular screws, the Iso-C3D and CT were accurate in all cases (sensitivities = 100%), and fluoroscopy was accurate in five of the eight trials for both joints (sensitivities = 63%). Finally, with screws protruding through the medial wall, the sensitivity of fluoroscopy was 25%, for the Iso-C3D 88% ( p = 0.02), and for CT was 100% ( p = 0.003). Conclusions: Three-dimensional fluoroscopic imaging is more accurate than 2D fluoroscopy for detection of intra-articular incongruities and implant position and is similar to CT. Clinical Relevance: This new technology may be particularly useful in assessing calcaneal fractures and may lead to improved fracture reduction, less implant misplacement, and improved patient outcomes.Keywords
This publication has 36 references indexed in Scilit:
- Calkaneusfrakturen - offene Reposition und interne StabilisierungZentralblatt fur Chirurgie, 2003
- Untersuchungen zur Bildqualität, Hochkontrastauflösung und Dosis am Stamm- und Gliedmaßenskelett mit einem neuen dedizierten CT-System (ISO-C-3D)RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2002
- Frakturdiagnostik am Kniegelenk mit einem neuen mobilen CT-System (ISO-C-3D): Vergleich mit konventionellem Röntgen und Spiral-CTRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2002
- Open Reduction and Internal Fixation of Type II Intra-Articular Calcaneus FracturesFoot & Ankle International, 1996
- Operative vs. Nonoperative Treatment of Intra-Articular Fractures of the Calcaneus: A Prospective Randomized TrialFoot & Ankle International, 1996
- Operative Treatment of Acute Displaced Intra-Articular Calcaneus FracturesFoot & Ankle International, 1995
- Extensive Intraarticular Fractures of the FootClinical Orthopaedics and Related Research, 1993
- The Operative Treatment of Intraarticular Calcaneal FracturesClinical Orthopaedics and Related Research, 1993
- Intraarticular fractures of the calcaneus: value of CT findings in determining prognosis.American Journal of Roentgenology, 1992
- Roentgen Examination of the Subtaloid Joint in Fractures of the CalcaneusActa Radiologica, 1949