Effect of lung compliance and endotracheal tube leakage on measurement of tidal volume
Open Access
- 6 October 2004
- journal article
- research article
- Published by Springer Nature in Critical Care
- Vol. 8 (6) , R398-402
Abstract
Introduction: The objective of this laboratory study was to measure the effect of decreased lung compliance and endotracheal tube (ETT) leakage on measured exhaled tidal volume at the airway and at the ventilator, in a research study with a test lung. Methods: The subjects were infant, adult and pediatric test lungs. In the test lung model, lung compliances were set to normal and to levels seen in acute respiratory distress syndrome. Set tidal volume was 6 ml/kg across a range of simulated weights and ETT sizes. Data were recorded from both the ventilator light-emitting diode display and the CO2SMO Plus monitor display by a single observer. Effective tidal volume was calculated from a standard equation. Results: In all test lung models, exhaled tidal volume measured at the airway decreased markedly with decreasing lung compliance, but measurement at the ventilator showed minimal change. In the absence of a simulated ETT leak, calculation of the effective tidal volume led to measurements very similar to exhaled tidal volume measured at the ETT. With a simulated ETT tube leak, the effective tidal volume markedly overestimated tidal volume measured at the airway. Conclusion: Previous investigators have emphasized the need to measure tidal volume at the ETT for all children. When ETT leakage is minimal, it seems from our simulated lung models that calculation of effective tidal volume would give similar readings to tidal volume measured at the airway, even in small patients. Future studies of tidal volume measurement accuracy in mechanically ventilated children should control for the degree of ETT leakage.Keywords
This publication has 6 references indexed in Scilit:
- Accuracy of displayed values of tidal volume in the pediatric intensive care unit*Critical Care Medicine, 2002
- The influence of endotracheal tube leak on the assessment of respiratory function in ventilated childrenIntensive Care Medicine, 2001
- Implementation of a Low Tidal Volume Ventilation Protocol for Patients with Acute Lung Injury or Acute Respiratory Distress SyndromeRespiratory Care, 2001
- Tidal Volumes For Ventilated Infants Should Be Determined with a Pneumotachometer Placed at the Endotracheal TubeAmerican Journal of Respiratory and Critical Care Medicine, 2000
- Ventilation with Lower Tidal Volumes as Compared with Traditional Tidal Volumes for Acute Lung Injury and the Acute Respiratory Distress SyndromeNew England Journal of Medicine, 2000
- Cuffed endotracheal tubes in pediatric intensive careThe Journal of Pediatrics, 1994