Thin CdTe detector in diagnostic x-ray spectroscopy
- 17 April 2003
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 30 (5) , 771-777
Abstract
A CdTe Schottky diode detector of 1 mm thickness was employed in diagnostic x-ray spectroscopy. The detector response to monoenergetic photons was investigated with gamma rays from the calibration sources (241Am and 133Ba). As spectral distortion due to carrier trapping, known as tailing, was small in gamma-ray spectra, the effects of carrier trapping were not taken into account in the calculation of response functions. The distortion due to the transmission of primary x rays and the escape of secondary x rays (K-fluorescent x rays and Compton-scattered x rays) from the crystal was included in the calculated response functions. X-ray spectra corrected using the response functions were in good agreement with the reference spectra obtained with a high-purity germanium detector. The results indicated that correction for the distortion due to carrier trapping is not necessary when using a thin CdTe detector in diagnostic x-ray spectroscopy.Keywords
This publication has 15 references indexed in Scilit:
- An alignment method for mammographic X-ray spectroscopy under clinical conditionsBritish Journal of Radiology, 2002
- CdZnTe detector in diagnostic x‐ray spectroscopyMedical Physics, 2002
- High-resolution Schottky CdTe diode for hard X-ray and gamma-ray astronomyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1999
- Performance of a new Schottky CdTe detector for hard X-ray spectroscopyIEEE Transactions on Nuclear Science, 1998
- Evaluation of CdZnTe detectors for soft X-Ray applicationsIEEE Transactions on Nuclear Science, 1994
- Measurement of X-ray spectrum using a small size CdTe multichannel detectorIEEE Transactions on Nuclear Science, 1992
- The use of cadmium telluride detectors for the qualitative analysis of diagnostic x-ray spectraPhysics in Medicine & Biology, 1984
- Stripping of X-ray bremsstrahlung spectra up to 300 kVp on a desk type computerPhysics in Medicine & Biology, 1979
- The effects of carrier trapping in semiconductor gamma-ray spectrometersNuclear Instruments and Methods, 1969
- Zum Mechanismus des lichtelektrischen Prim rstromes in isolierenden KristallenThe European Physical Journal A, 1932