Interlaboratory reproducibility of mobility parameters in capillary electrophoresis for substance identification in systematic toxicological analysis
- 21 January 2002
- journal article
- Published by Wiley in Electrophoresis
- Vol. 23 (1) , 67-73
Abstract
An interlaboratory pilot study was performed to determine the reproducibility of mobility parameters in capillary zone electrophoresis (CZE) and micellar electrokinetic chromatography (MEKC). The study was performed by an intended small number of laboratories (three) that used different brands of instruments (two). The effective mobility was corrected using standards by a method that was recently introduced to obtain a more reproducible migration parameter. A test set of 20 acidic test compounds and 5 reference compounds were analyzed during five days in each laboratory using CZE and MEKC. Buffers used consisted of 90 mM borate set at pH 8.4 (CZE) and 20 mM phosphate, 50 mM sodium dodecyl sulfate set at pH 7.5 (MEKC). Analyses were carried out using fused-silica capillaries at an electric field strength of either 52.6 kV/m or 37.5 kV/m. The interlaboratory reproducibility (mean RSD) of the effective mobility was 3.0% for CZE and 6.7% for MEKC. After applying the correction method, these values became 3.0% for CZE and 3.3% for MEKC, which is adequate for systematic toxicological analysis (STA) applications. A significant improvement of reproducibility for the calculated corrected effective mobility mu(eff)c was observed when variations are high. Therefore, it is recommended to use the correction method in interlaboratory situations, especially when instruments and capillaries from different manufacturers are used.Keywords
This publication has 15 references indexed in Scilit:
- Determination of morphine analogues, caffeine and amphetamine in biological fluids by capillary electrophoresis with the marker techniqueJournal of Chromatography A, 1996
- Marker techniques for high-accuracy identification in CZEAnalytical Chemistry, 1995
- Inter-company cross validation exercise on capillary electrophoresis. Quantitative determination of drug counter-ion levelChromatographia, 1995
- An inter-company cross-validation exercise on capillary electrophoresis testing of dose uniformity of paracetamol content in formulationsChromatographia, 1994
- Inter-company cross-validation exercise on capillary electrophoresis: I. Chiral analysis of clenbuterolJournal of Chromatography A, 1993
- An overview on the standardization of chromatographic methods for screening analysis in toxicology by means of retention indices and secondary standardsAnalytical and Bioanalytical Chemistry, 1993
- An overview on the standardization of chromatographic methods for screening analysis in toxicology by means of retention indices and secondary standardsAnalytical and Bioanalytical Chemistry, 1993
- Correction of retention index values in high-performance liquid chromatography as a tool for comparison of results obtained with different octadecyl silica phasesJournal of Chromatography A, 1987
- Application of RF correction in thin-layer chromatography by means of two reference RF valuesJournal of Chromatography A, 1972
- Application of RF correction in thin-layer chromatography by means of two reference RF valuesJournal of Chromatography A, 1970