Comparison of two portable solid state detectors with an improved collimation and alignment device for mammographic x-ray spectroscopy
- 30 August 2006
- journal article
- Published by Wiley in Medical Physics
- Vol. 33 (9) , 3469-3477
Abstract
We describe a portable system for mammographic x-ray spectroscopy, based on a 2 × 2 × 1 mm 3 cadmium telluride ( CdTe ) solid state detector, that is greatly improved over a similar system based on a 3 × 3 × 2 mm 3 cadmium zinc telluride (CZT) solid state detector evaluated in an earlier work. The CdTe system utilized new pinhole collimators and an alignment device that facilitated measurement of mammographicx-ray spectra.Mammographicx-ray spectra acquired by each system were comparable. Half value layer measurements obtained using an ion chamber agreed closely with those derived from the x-ray spectra measured by either detector. The faster electronics and other features of the CdTe detector allowed its use with a larger pinhole collimator than could be used with the CZT detector. Additionally, the improved pinhole collimator and alignment features of the apparatus permitted much more rapid setup for acquisition of x-ray spectra than was possible on the system described in the earlier work. These improvements in detector technology, collimation and ease of alignment, as well as low cost, make this apparatus attractive as a tool for both laboratory research and advanced mammography quality control.Keywords
This publication has 6 references indexed in Scilit:
- Direct analysis of molybdenum target generated x‐ray spectra with a portable deviceMedical Physics, 2004
- CdZnTe detector in mammographic x-ray spectroscopyPhysics in Medicine & Biology, 2002
- CdZnTe detector in diagnostic x‐ray spectroscopyMedical Physics, 2002
- An accurate method for computer‐generating tungsten anode x‐ray spectra from 30 to 140 kVMedical Physics, 1997
- On the measurement of half‐value layer in film–screen mammographyMedical Physics, 1990
- An Algorithm for Least-Squares Estimation of Nonlinear ParametersJournal of the Society for Industrial and Applied Mathematics, 1963