As a flexible needle with a bevel tip is pushed through soft tissue, the asymmetry of the tip causes the needle to bend. We propose that, by using nonholonomic kinematics, control, and path planning, an appropriately designed needle can be steered through tissue to reach a specified 3D target. Such steering capability could enhance targeting accuracy and may improve outcomes for percutaneous therapies, facilitate research on therapy effectiveness, and eventually enable new minimally invasive techniques. In this paper, we consider a first step toward active needle steering: design and experimental validation of a nonholonomic model for steering flexible needles with bevel tips. The model generalizes the standard three degree-of-freedom (DOF) nonholonomic unicycle and bicycle models to 6 DOF using Lie group theory. Model parameters are fit using experimental data, acquired via a robotic device designed for the specific purpose of inserting and steering a flexible needle. The experiments quantitatively validate the bevel-tip needle steering model, enabling future research in flexible needle path planning, control, and simulation.Keywords
This publication has 39 references indexed in Scilit:
- A phase I clinical study of biweekly pemetrexed and gemcitabine in patients with advanced solid tumorsJournal of Clinical Oncology, 2004
- Cancer Statistics, 2004CA: A Cancer Journal for Clinicians, 2004
- Needle insertion modeling and simulationIEEE Transactions on Robotics and Automation, 2003
- Needle Steering and Model-Based Trajectory Planning2003
- A Model for Relations Between Needle Deflection, Force, and Thickness on Needle PenetrationLecture Notes in Computer Science, 2001
- Robotic System for Biopsy and Therapy of Breast Lesions in a High-Field Whole-Body Magnetic Resonance Tomography UnitInvestigative Radiology, 2000
- Schemes for the identification of tissue types and boundaries at the tool point for surgical needlesIEEE Transactions on Information Technology in Biomedicine, 2000
- Elastic Moduli of Breast and Prostate Tissues under CompressionUltrasonic Imaging, 1998
- Simulation of resistance forces acting on surgical needlesProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1997