Abstract
A frequency‐stabilized single‐mode cw 118.8‐μm CH3OH waveguide laser was newly developed for polari‐interferometry on the JT‐60 tokamak. Interferometric plasma diagnostics on large fusion devices requires a mode purity of a source laser for the long‐distance propagation and frequency stability to obtain an adequate fringe resolution. The accomplished performances are Δf/f−8, power stability 11 mode to yield a Gaussian‐shaped profile in the far field. Homogeneous pumping of laser media with a Ge étalon beam coupler and feedback stability control are the keys to the source far‐infrared laser performances required for large tokamak diagnostics. The developed laser was totally adjustment free for more than 2 weeks and reliably in operation for 2 years, including the real‐time electron density feed‐back control on the JT‐60 tokamak.