Improving the accuracy of the transient plane source method by correcting probe heat capacity and resistance influences
Authors:
- Yanning LiYLYanning Li
- Chunfeng ShiCSChunfeng Shi
- Jian LiuJLJian Liu
- Errui LiuELErrui Liu
- Jian ShaoJSJian Shao
- Zhi ChenZCZhi Chen
- Dante J Dorantes-GonzalezDDDante J Dorantes-Gonzalez
- Xiaotang HuXHXiaotang Hu
- 4 December 2013
- journal article
- Published by IOP Publishing in Measurement Science and Technology
- Vol. 25 (1) , 015006
Abstract
The transient plane source (TPS) method is a relatively newly developed transient approach for thermal conductivity measurement. Compared with the steady-state method, it is fast, and applicable to either solid, liquid or gas state materials; therefore, it has gained much popularity in recent years. However, during measurement, the measured power is influenced by the heat capacity of the electrical isolation films as well as the electrical resistance change of the metallic thin wire of the TPS probes. This further influences the measurement precision. Meanwhile, these two factors have been ignored in the traditional model of TPS developed by Gustafsson. In this paper, the influence of both the heat capacity and the resistance change of the TPS probe on the measured power is studied, and mathematical formulas relating the two factors and their respective corrections are deduced. Thereafter an improved model is suggested based on the traditional TPS model and the above theoretical models. Experiments on polymethylmethacrylate (PMMA) standard materials have been conducted using a home-made system, including TPS probes, data acquisition module and analysis software. The results show that the improved model can effectively improve the measurement precision of the TPS method by about 1.8–2.3% as evaluated by relative standard deviation.Keywords
This publication has 23 references indexed in Scilit:
- Analysis of the temperature distribution in a guarded hot plate apparatus for measuring thermal conductivityApplied Thermal Engineering, 2009
- The use of the transient hot-wire technique for measurement of the thermal conductivity of an epoxy-resin reinforced with glass fibres and/or carbon multi-walled nanotubesComposites Science and Technology, 2008
- An experimental study on the thermal conductivity of aluminium foams by using the transient plane source methodInternational Journal of Heat and Mass Transfer, 2008
- Thermal conductivity measurement and sedimentation detection of aluminum oxide nanofluids by using the 3ω methodInternational Journal of Heat and Fluid Flow, 2008
- Thermal conductivity of a metal-organic framework (MOF-5): Part II. MeasurementInternational Journal of Heat and Mass Transfer, 2007
- A new method for measuring thermal conductivity of thin filmsReview of Scientific Instruments, 1997
- Thermal conductivity ofa-Si:H thin filmsPhysical Review B, 1994
- Thermal Analysis of Materials1993
- Thermal conductivity measurement from 30 to 750 K: the 3ω methodReview of Scientific Instruments, 1990
- Flash Method of Determining Thermal Diffusivity, Heat Capacity, and Thermal ConductivityJournal of Applied Physics, 1961