A Pilot Study of Left Tracheal Pulse Oximetry
Authors:
- J. BrimacombeJBJ. Brimacombe
- C. KellerCKC. Keller
- J. MargreiterJMJ. Margreiter
- 2 October 2000
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 91 (4) , 1003-1006
Abstract
We tested the hypothesis that left tracheal pulse oximetry (Spo2) is more accurate than finger Spo2 when compared with oxygen saturation from arterial blood samples (Sao2) in anesthetized patients with normal thoracic anatomy. We also tested the hypothesis that tracheal oximetry readings are primarily derived from the tracheal mucosa. We studied 20 hemodynamically stable, well oxygenated, anesthetized patients with normal anatomy (ASA physical status I–III, 18–80 yr old). A single-use pediatric pulse oximeter was attached to the left lateral surface of a tracheal tube cuff. Tracheal and finger Spo2 (dominant index finger), and Sao2 (nondominant radial artery) were taken with the intracuff pressure at 0–60 cm H2O. Tracheal Spo2 was the same as Sao2 at an intracuff pressure of 10–60 cm H2O, but was less when the intracuff pressure was zero (P < 0.0001). Tracheal Spo2 was higher than finger Spo2 at an intracuff pressure of 10–60 cm H2O (all:P < 0.001), but was lower when the intracuff pressure was zero (P < 0.0001). Sao2 was always higher than finger Sao2 (P < 0.0001). Tracheal Spo2 was lower at an intracuff pressure of zero (P < 0.0001), but was otherwise similar over the range of intracuff pressures. Sao2 and finger Spo2 did not vary with intracuff pressure. Tracheal Spo2 agrees more closely with Sao2 than finger Spo2 at an intracuff pressure of 10–60 cm H2O (mean difference < 0.2%). We conclude that left tracheal Spo2 is feasible and provides similar readings to arterial blood samples and more accurate readings than finger oximetry in hemodynamically stable, well oxygenated, anesthetized patients with normal thoracic anatomy. Tracheal oximetry readings are not primarily derived from the tracheal mucosa. The technique merits further evaluation. Tracheal oximetry is feasible and provides similar readings to arterial blood samples and more accurate readings than finger oximetry in hemodynamically stable, well oxygenated, anesthetized patients with normal thoracic anatomy. Tracheal oximetry readings are not primarily derived from the tracheal mucosa. The technique merits further evaluation.Keywords
This publication has 8 references indexed in Scilit:
- A Pilot Study of Pharyngeal Pulse Oximetry with the Laryngeal Mask Airway: A Comparison with Finger Oximetry and Arterial Saturation Measurements in Healthy Anesthetized PatientsAnesthesia & Analgesia, 2000
- Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adultsHeart & Lung, 1998
- Buccal Pulse Oximeter Is More Accurate Than Finger Pulse Oximeter in Measuring Oxygen SaturationAnesthesia & Analgesia, 1992
- Pulse oximetry from the nasal septumJournal of Clinical Anesthesia, 1991
- Monitoring of Arterial Hemoglobin Oxygen Saturation Using a Tongue SensorAnesthesia & Analgesia, 1988
- MORPHINE CONJUGATES AND VENTILATORY DEPRESSION IN THE DOGAnesthesia & Analgesia, 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs.BMJ, 1984