Mycobacterium chimaera infections among cardiothoracic surgery patients associated with heater-cooler devices—Kansas and California, 2019
Authors:
- 6 October 2021
- journal article
- research article
- Published by Cambridge University Press (CUP) in Infection Control & Hospital Epidemiology
- Vol. 43 (10) , 1333-1338
Abstract
Background: In 2015, an international outbreak of Mycobacterium chimaera infections among patients undergoing cardiothoracic surgeries was associated with exposure to contaminated LivaNova 3T heater-cooler devices (HCDs). From June 2017 to October 2020, the Centers for Disease Control and Prevention was notified of 18 patients with M. chimaera infections who had undergone cardiothoracic surgeries at 2 hospitals in Kansas (14 patients) and California (4 patients); 17 had exposure to 3T HCDs. Whole-genome sequencing of the clinical and environmental isolates matched the global outbreak strain identified in 2015.Methods: Investigations were conducted at each hospital to determine the cause of ongoing infections. Investigative methods included query of microbiologic records to identify additional cases, medical chart review, observations of operating room setup, HCD use and maintenance practices, and collection of HCD and environmental samples.Results: Onsite observations identified deviations in the positioning and maintenance of the 3T HCDs from the US Food and Drug Administration (FDA) recommendations and the manufacturer’s updated cleaning and disinfection protocols. Additionally, most 3T HCDs had not undergone the recommended vacuum and sealing upgrades by the manufacturer to decrease the dispersal of M. chimaera–containing aerosols into the operating room, despite hospital requests to the manufacturer.Conclusions: These findings highlight the need for continued awareness of the risk of M. chimaera infections associated with 3T HCDs, even if the devices are newly manufactured. Hospitals should maintain vigilance in adhering to FDA recommendations and the manufacturer’s protocols and in identifying patients with potential M. chimaera infections with exposure to these devices.Keywords
This publication has 8 references indexed in Scilit:
- Notes from the Field: Mycobacteria chimaera Infections Associated with Heater-Cooler Unit Use During Cardiopulmonary Bypass Surgery — Los Angeles County, 2012–2016MMWR. Morbidity and Mortality Weekly Report, 2019
- Complete Genome Sequence of Mycobacterium chimaera Strain CDC2015-22-71Microbiology Resource Announcements, 2017
- Invasive Nontuberculous Mycobacterial Infections among Cardiothoracic Surgical Patients Exposed to Heater–Cooler Devices1Emerging Infectious Diseases, 2017
- Clinical implications of Mycobacterium chimaera detection in thermoregulatory devices used for extracorporeal membrane oxygenation (ECMO), Germany, 2015 to 2016Eurosurveillance, 2016
- Notes from the Field:Mycobacterium chimaeraContamination of Heater-Cooler Devices Used in Cardiac Surgery—United StatesMMWR. Morbidity and Mortality Weekly Report, 2016
- Decontamination of heater–cooler units associated with contamination by atypical mycobacteriaJournal of Hospital Infection, 2016
- Contamination during production of heater-cooler units by Mycobacterium chimaera potential cause for invasive cardiovascular infections: results of an outbreak investigation in Germany, April 2015 to February 2016Eurosurveillance, 2016
- Prolonged Outbreak of Mycobacterium chimaera Infection After Open-Chest Heart SurgeryClinical Infectious Diseases, 2015