Non-invasive temperature monitoring and hyperthermic injury onset detection using X-ray CT during HIFU thermal treatment in ex vivo fatty tissue
Authors:
- 17 February 2014
- journal article
- Published by Taylor & Francis in International Journal of Hyperthermia
- Vol. 30 (2) , 119-125
Abstract
This paper examines X-ray CT, to serve as an image-guiding thermal monitoring modality for high intensity focused ultrasound (HIFU) treatment of fatty tissues. Six ex vivo porcine fat tissue specimens were scanned by X-ray CT simultaneously with the application of HIFU. Images were acquired during both heating and post-ablation stages. The temperature at the focal zone was measured simultaneously using a thermocouple. The mean values of the Hounsfield units (HU) at the focal zone were registered and plotted as a function of temperature. In all specimens studied, the HU versus temperature curves measured during the heating stage depicted a characteristic non-linear parabolic trajectory (R(2) > 0.87). The HU-temperature trajectory initially decreased to a minimum value at about 44.5 °C and then increased substantially as the heating progressed. The occurrence of this nadir point during the heating stage was clearly detectable. During post-ablation cooling, on the other hand, the HU increased monotonically with the decreasing temperature and depicted a clearly linear trajectory (R(2) ≥ 0.9). Our results demonstrate that the HU-temperature curve during HIFU treatment has a characteristic parabolic trajectory for fat tissue that might potentially be utilised for thermal monitoring during HIFU ablation treatments. The clear detection of 44.5 °C, presumably marking the onset of hyperthermic injury, can be detected non-invasively as an occurrence of a minimum on the HU-time curve without any need to relate the HU directly to temperature. Such features may be helpful in monitoring and optimising HIFU thermal treatment for clinically applicable indications such as in the breast by providing a non-invasive monitoring of tissue damage.Keywords
This publication has 26 references indexed in Scilit:
- HIFU treatment time reduction in superficial tumours through focal zone path selectionInternational Journal of Hyperthermia, 2011
- Feasibility of Noninvasive Temperature Assessment During Radiofrequency Liver Ablation on Computed TomographyJournal of Computer Assisted Tomography, 2011
- Feasibility of computed tomography based thermometry during interstitial laser heating in bovine liverEuropean Radiology, 2011
- Temperature‐induced tissue susceptibility changes lead to significant temperature errors in PRFS‐based MR thermometry during thermal interventionsMagnetic Resonance in Medicine, 2010
- Hyperthermic injury to adipocyte cells by selective heating of subcutaneous fat with a novel radiofrequency device: Feasibility studiesLasers in Surgery and Medicine, 2010
- Noninvasive temperature mapping with MRI using chemical shift water‐fat separationMagnetic Resonance in Medicine, 2010
- Multi-slice computed tomography: A tool for non-invasive temperature measurement?International Journal of Hyperthermia, 2010
- MR thermometryJournal of Magnetic Resonance Imaging, 2008
- High intensity focused ultrasound: surgery of the future?British Journal of Radiology, 2003
- MRI evaluation of thermal ablation of tumors with focused ultrasoundJournal of Magnetic Resonance Imaging, 1998