Indications and immediate and long‐term results of a novel pericardium covered stent graft: Consecutive 5 year single center experience
- 6 November 2015
- journal article
- research article
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 87 (4) , 712-719
Abstract
Background: The use of covered stent grafts during percutaneous coronary intervention (PCI) is a life saving solution to seal acute iatrogenic vessel rupture. However, the presence of an impenetrable mechanical barrier is also appealing during treatment of friable coronary plaques but the synthetic PTFE‐membrane that might trigger excessive neointimal proliferation has limited its elective‐use. Pericardium tissue may offer an appealing “natural” alternative. Aim of our study is to report the consecutive 5‐year single center experience with the use of pericardium‐covered stents (PCS) (ITGI‐Medical, Israel) in a variety of emergency and elective applications.Methods: Nineteen consecutive patients undergoing implantation of PCS at the Royal Brompton in the last 5‐years. Reasons for PCS implantation included treatment of degenerated vein grafts, large coronary aneurysms, and acute iatrogenic vessel rupture.Results: Angiographic success, defined as the ability of the device to be deployed in the indexed lesion with no contrast extravasation with residual angiographic stenosis <30% and a final thrombolysis in myocardial infarction (TIMI)‐3 flow was achieved in all cases. Procedural success, defined as the achievement of angiographic success without any major adverse cardiovascular event (MACE) was achieved in 94.7% of patients. In‐stent restenosis (ISR) was observed in 26.3% and all patients underwent successful target vessel revascularization with DES (mean time to restenosis 9.0 ± 4.0 months). At a mean follow‐up of 32.5 ± 23.3 months no acute or late stent thrombosis was observed.Conclusion: PCSs were effective in the treatment of friable embolization‐prone coronary plaques, sealing of acute iatrogenic vessel rupture and exclusion of large aneurysms with no thrombosis but high target lesion revascularization.Keywords
Funding Information
- The Scientific and Technological Research Council of Turkey (TUBITAK)
This publication has 25 references indexed in Scilit:
- The pericardium covered stent (PCS)EuroIntervention, 2009
- A randomized trial of polytetrafluoroethylene-membrane-covered stents compared with conventional stents in aortocoronary saphenous vein graftsJACC, 2003
- Randomized Evaluation of Polytetrafluoroethylene-Covered Stent in Saphenous Vein GraftsCirculation, 2003
- Comparison of PTFE, pericardium bovine and fascia lata for repair of incisional hernia in rat model, experimental studyHernia, 2002
- Bovine Pericardium vs Dacron for Patch Angioplasty After Carotid EndarterectomyArchives of Surgery, 2002
- Membrane-covered stents: A new treatment strategy for saphenous vein graft lesionsCatheterization and Cardiovascular Interventions, 2001
- Polytetrafluoroethylene-covered stent for the treatment of narrowings in aorticocoronary saphenous vein graftsThe American Journal of Cardiology, 2000
- Coronary stent grafts covered by a polytetrafluoroethylene membraneThe American Journal of Cardiology, 1999
- Delayed development of a pseudoaneurysm in the left circumflex artery following angioplasty and stent placement, treated with intravascular ultrasound-guided stentingCatheterization and Cardiovascular Diagnosis, 1997
- Obliteration of a coronary artery aneurysm with percutaneous transluminal coronary angioplasty and stent placementCatheterization and Cardiovascular Diagnosis, 1997