Precise frequency measurements of atomic transitions by use of a Rb-stabilized resonator
- 1 September 2003
- journal article
- Published by Optica Publishing Group in Optics Letters
- Vol. 28 (17) , 1579-1581
Abstract
We demonstrate a technique for frequency measurements of atomic transitions with a precision of 30 kHz. The frequency is measured with a ring-cavity resonator whose length is calibrated against a reference laser locked to the line of , the frequency of which is known with 10-kHz accuracy. We have used this method to measure the hyperfine structure in the state of . We obtain precise values for the hyperfine constants, and , and a value of 77.992(20) MHz for the isotope shift in the line.
Keywords
All Related Versions
This publication has 9 references indexed in Scilit:
- High-accuracy wavemeter based on a stabilized diode laserApplied Physics Letters, 2001
- Absolute Optical Frequency Measurement of the CesiumLine with a Mode-Locked LaserPhysical Review Letters, 1999
- Measurement of Parity Nonconservation and an Anapole Moment in CesiumScience, 1997
- Hyperfine structure and absolute frequency of the ^87Rb 5P_3/2 stateOptics Letters, 1996
- Correction to the Updated Edlén Equation for the Refractive Index of AirMetrologia, 1994
- Precision measurement of the photon recoil of an atom using atomic interferometryPhysical Review Letters, 1993
- An Updated Edlén Equation for the Refractive Index of AirMetrologia, 1993
- Frequency measurements on optically narrowed Rb-stabilised laser diodes at 780 nm and 795 nmApplied Physics B, 1991
- Experimental determinations of the hyperfine structure in the alkali atomsReviews of Modern Physics, 1977