Intensity‐modulated radiation therapy for multiple targets with tomotherapy using multiple sets of static ports from different angles
Open Access
- 6 April 2020
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 21 (6) , 132-138
Abstract
Background To treat multiple targets separated in the craniocaudal direction within a short time, we invented a new technique using multiple static‐port tomotherapy with the dynamic‐jaw mode and named it the pseudo‐DJDC (pDJDC) technique. We compared the pDJDC plans and helical tomotherapy plans using the dynamic‐jaw mode (HDJ) for multiple targets. In the pDJDC plans, we used a beam set with 2–7 ports to the targets at the same level in the craniocaudal direction, and employed another beam set for other targets using different port angles (9–12 angles in total). Methods In seven patients, two plans using the pDJDC and HDJ techniques were compared. For multiple targets (n = 2–6), 20–60 Gy in 2‐ to 7.5‐Gy fractions were prescribed for the planning target volumes at D50%. The conformity index, uniformity index (D5%/D95%), dose distribution in the lung, and treatment time were evaluated. Results The median conformity index of all seven patients was 3.0 for the pDJDC plans and 2.4 for the HDJ plans (P = 0.031). The median uniformity indices of the planning target volume (n = 25) for the two plans were 1.048 and 1.057, respectively (P = 0.10). For five patients with thoracic targets, the median mean lung doses were 2.6 Gy and 2.4 Gy, respectively (P = 0.63). The median V5Gy and V20Gy of the lungs in the five patients were 11.8% and 8.5% (P = 0.63), and 1.6% and 2.1% (P = 0.31), respectively. The pDJDC plans reduced the treatment time by 48% compared to the HDJ plans (median: 462 and 884 sec, respectively, P = 0.031). Conclusion The pDJDC technique allows treatment of multiple targets in almost half the time of the HDJ technique. The pDJDC plans were comparable to the HDJ plans in dose distribution, although the conformity index deteriorated.Keywords
Funding Information
- Japan Society for the Promotion of Science (JP18K15285)
This publication has 16 references indexed in Scilit:
- Efficacy of stereotactic radiotherapy for brain metastases using dynamic jaws technology in the helical tomotherapy systemBritish Journal of Radiology, 2016
- Benchmark Credentialing Results for NRG-BR001: The First National Cancer Institute-Sponsored Trial of Stereotactic Body Radiation Therapy for Multiple MetastasesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2016
- Efficacy of the Dynamic Jaw Mode in Helical Tomotherapy With Static Ports for Breast CancerTechnology in Cancer Research & Treatment, 2014
- Helical and Static-port Tomotherapy Using the Newly-developed Dynamic Jaws Technology for Lung CancerTechnology in Cancer Research & Treatment, 2014
- Intensity-modulated radiation therapy using static ports of tomotherapy (TomoDirect): comparison with the TomoHelical modeRadiation Oncology, 2013
- Accelerated large volume irradiation with dynamic Jaw/Dynamic Couch Helical TomotherapyRadiation Oncology, 2012
- Der ICRU-Report 83: Verordnung, Dokumentation und Kommunikation der fluenzmodulierten Photonenstrahlentherapie (IMRT)Strahlentherapie und Onkologie, 2012
- History of tomotherapyPhysics in Medicine & Biology, 2006
- Evaluation of two tomotherapy-based techniques for the delivery of whole-breast intensity-modulated radiation therapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Tomotherapy: A new concept for the delivery of dynamic conformal radiotherapyMedical Physics, 1993