Ultra‐high performance supercritical fluid chromatography coupled to tandem mass spectrometry for antidoping analyses: Assessment of the inter‐laboratory reproducibility with urine samples
Open Access
- 5 December 2020
- journal article
- research article
- Published by Wiley in Analytical Science Advances
- Vol. 2 (1-2) , 68-75
Abstract
The aim of this study was to assess the interlaboratory reproducibility of ultra‐high performance supercritical fluid chromatography coupled with tandem mass spectrometry method for routine antidoping analyses. To do so, a set of 21 doping agents, spiked in urine and analyzed after dilute and shoot treatment, was used to assess the variability of their retention times between four different laboratories, all equipped with the same chromatographic system and with the same ultra‐high performance supercritical fluid chromatography stationary phase chemistry. The average relative standard deviations (RSD%) demonstrated a good reproducibility of the retention times for 19 out of 21 analytes, with RSD% values below 3.0%. Only for two substances, namely fenbutrazate and niketamide, the retention was not repeatable between laboratories, with RSD% of approximately 15% in both cases. This behaviour was associated with (a) the low organic modifier percentage (around 2‐4%) in the mobile phase at the corresponding retention times, and (b) the influence of the system volume on poorly retained analytes. An analysis on seven “blind” urines was subsequently carried out in the same four laboratories. In these blind samples, either one, two, or none of the 21 doping agents previously analyzed were present at an unknown concentration. Each laboratory had to perform the identification of the compounds in the samples and estimate their concentrations. All laboratories assigned all target analytes correctly in all blind urine samples and provide a comparable estimation of their concentrations.Keywords
Funding Information
- World Anti-Doping Agency (T17R02TK)
- European Regional Development Fund (CZ.02.1.01/0.0/0.0/15_003/0000465)
This publication has 16 references indexed in Scilit:
- The utility of ultra-high performance supercritical fluid chromatography–tandem mass spectrometry (UHPSFC-MS/MS) for clinically relevant steroid analysisJournal of Chromatography B, 2018
- Enantiomeric separation and quantification of R/S-amphetamine in urine by ultra-high performance supercritical fluid chromatography tandem mass spectrometryJournal of Chromatography B, 2018
- High resolution full scan liquid chromatography mass spectrometry comprehensive screening in sports antidoping urine analysisJournal of Pharmaceutical and Biomedical Analysis, 2018
- Fast and sensitive supercritical fluid chromatography – tandem mass spectrometry multi-class screening method for the determination of doping agents in urineAnalytica Chimica Acta, 2016
- Analytical Strategies for Doping Control Purposes: Needs, Challenges, and PerspectivesAnalytical Chemistry, 2015
- The many faces of packed column supercritical fluid chromatography – A critical reviewJournal of Chromatography A, 2015
- Ultra high performance supercritical fluid chromatography coupled with tandem mass spectrometry for screening of doping agents. II: Analysis of biological samplesAnalytica Chimica Acta, 2015
- Maximizing kinetic performance in supercritical fluid chromatography using state-of-the-art instrumentsJournal of Chromatography A, 2013
- Retention Behavior of Triglycerides in Octadecyl Packed Subcritical Fluid Chromatography with CO2/Modifier Mobile PhasesAnalytical Chemistry, 1999
- Effect of temperature and mobile phase density on retention and efficiency in supercritical fluid chromatographyJournal of High Resolution Chromatography, 1987