Innovative technologies for the assessment of cardiovascular medical devices: state-of-the-art techniques for artificial heart valve testing
Authors:
- Mauro GrigioniMGMauro Grigioni
- Carla DanieleCDCarla Daniele
- Giuseppe D'AvenioGDGiuseppe D'Avenio
- Umberto MorbiducciUmberto Morbiducci
- Costantino Del GaudioCGCostantino Del Gaudio
- Mara AbbateMAMara Abbate
- David Di MeoDMDavid Di Meo
- 1 September 2004
- journal article
- review article
- Published by Taylor & Francis in Expert Review of Medical Devices
- Vol. 1 (1) , 81-93
Abstract
Prosthetic heart valves (PHVs) are engineered devices used for replacing diseased natural cardiac valves. This article presents several investigational techniques for the evaluation of the performance of these clinical devices, whose implantation is not completely free of drawbacks. The state-of-the-art in the technological approach for PHV testing is addressed. As the fluid dynamics of PHVs are particularly complex, the main focus will be on experimental velocimetric techniques and computational analysis. A methodology for the analysis of the valve's signature, in terms of its characteristic sound in the opening and closing phases, is also presented. The aforementioned techniques are necessary to guarantee an operational life of the implanted device as free as possible from clinical complications. It can be realistically expected that this characterization will help designers in improving PHV performance.Keywords
This publication has 45 references indexed in Scilit:
- Lowering the Intensity of Oral Anticoagulant TherapyArchives of internal medicine (1960), 2004
- Laser Doppler Anemometry Study of Bidimensional Flows Downstream of Three 19 mm Bileaflet Valves in the Mitral Position, Under Kinematic SimilarityAnnals of Biomedical Engineering, 2000
- Determination of Principal Reynolds Stresses in Pulsatile Flows After Elliptical Filtering of Discrete Velocity MeasurementsJournal of Biomechanical Engineering, 1993
- Two-dimensional color-mapping of turbulent shear stress distribution downstream of two aortic bioprosthetic valves in vitroJournal of Biomechanics, 1992
- A Study of Prosthetic Heart Valve SoundsIEEE Transactions on Biomedical Engineering, 1987
- A 6 year experience with the St. Jude Medical Valve: Hemodynamic performance, surgical results, biocompatibility and follow-upJACC, 1985
- Human red blood cell hemolysis in a turbulent shear flow: Contribution of Reynolds shear stressesBiorheology, 1984
- Deformation and fragmentation of human red blood cells in turbulent shear flowBiophysical Journal, 1975
- Measured Turbulence and Its Effect on Thrombus FormationCirculation Research, 1974
- Statistical Errors in Measurements on Random Time FunctionsJournal of Applied Physics, 1952