Temporal response of medical liquid crystal displaysa)
Authors:
- 25 January 2007
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 34 (2) , 639-646
Abstract
Displays based on liquid crystal technology suffer from slow temporal response due to the dynamics of the molecular rearrangement in response to a pixel voltage change. A slow display can affect the visualization by the human observer of subtle contrast in dynamic presentation of volumetric image datasets or real‐time image sequences. In this paper, we describe a measurement method for the characterization of the temporal response of medical liquid crystal displays (LCDs). The ratio of luminance difference to noise at the gray levels of concern determines the reliability of measurements. Coefficients of variations are used to represent the measurement reliability. We optimized the repeatability of most response time measurements to less than 10%. However, poor repeatability is encountered for the response of adjacent gray levels. inter‐gray‐level transition time matrices were measured for four medical displays and one high‐definition TV LCD display. Response times range from below to above . For each display, response times are not uniformly distributed, with a faster response for large gray‐level transitions. Transition times are smaller when the starting gray level is between 10 and 20 for a target between 25 and 150. The difference could be over for different transitions within a display. For transitions with poor temporal response, the luminance after 1, 3, and 5 frames reaches only 12, 45, and 75% of the target value, respectively. We also found that LCD response time depends on temperature, with warm‐up reducing the response time by a factor of 2.Keywords
This publication has 8 references indexed in Scilit:
- Precision of gray level response time measurements of medical liquid crystal displayReview of Scientific Instruments, 2006
- 19.1: LCD Visual Quality Analysis by Moving Picture SimulationSID Symposium Digest of Technical Papers, 2005
- AAPM/RSNA Tutorial on Equipment Selection: PACS Equipment OverviewRadioGraphics, 2004
- 7.3: Motion ArtifactsSID Symposium Digest of Technical Papers, 2004
- Motion blur reduction for liquid crystal displays: motion-compensated inverse filtering2004
- AAPM/RSNA Tutorial on Equipment Selection: PACS Equipment OverviewRadioGraphics, 2004
- Optimized active-matrix drives for liquid crystal displaysJournal of Applied Physics, 2001
- TFT/LCD1996