Abstract
Purpose: To optimize and compare the performance of 3D UTE and zero echo time (ZTE) MRI sequences for high‐resolution lung imaging at 3 T. Methods: UTE and ZTE sequences were optimized for lung imaging through phantom studies, and in vivo experiments. Imaging parameters, including bandwidth (BW) and flip angle (FA), were systematically assessed to identify the optimal settings for both sequences. A novel transient UTE sequence was introduced to enhance the SNR by allowing magnetization recovery during the respiratory wait period. Numerical simulations and in vivo experiments were conducted to compare the SNR performance of the proposed transient UTE with conventional steady‐state UTE. Results: UTE imaging with a BW of ±125 kHz and an FA of 2° produced the sharpest lung images without noticeable artifacts. In contrast, ZTE imaging was optimal at a lower BW of ±62.5 kHz and an FA of 2°, but showed greater blurriness and more pronounced inhomogeneous excitation artifacts than UTE. The transient UTE sequence increased lung imaging SNR by approximately 11% to 20% compared to steady‐state UTE without introducing obvious artifacts. Conclusion: UTE outperformed ZTE in terms of image sharpness and artifact reduction for lung imaging. The proposed transient UTE sequence further enhanced SNR, demonstrating the potential for improved lung MRI quality and diagnostic accuracy.Keywords
Funding Information
- GE Healthcare
- National Institutes of Health (R01AR079484, RF1AG075717)
This publication has 39 references indexed in Scilit:
- Assessment of Lung Nodule Detection and Lung CT Screening Reporting and Data System Classification Using Zero Echo Time Pulmonary MRIJournal of Magnetic Resonance Imaging, 2024
- Balanced steady‐state free precession thoracic imaging with half‐radial dual‐echo readout on smoothly interleaved archimedean spiralsMagnetic Resonance in Medicine, 2019
- Iterative motion‐compensation reconstruction ultra‐short TE (iMoCo UTE) for high‐resolution free‐breathing pulmonary MRIMagnetic Resonance in Medicine, 2019
- Motion robust high resolution 3D free‐breathing pulmonary MRI using dynamic 3D image self‐navigatorMagnetic Resonance in Medicine, 2017
- Three‐dimensional T1 and T2* mapping of human lung parenchyma using interleaved saturation recovery with dual echo ultrashort echo time imaging (ITSR‐DUTE)Journal of Magnetic Resonance Imaging, 2016
- UTE imaging with simultaneous water and fat signal suppression using a time-efficient multispoke inversion recovery pulse sequenceMagnetic Resonance in Medicine, 2015
- ZTE imaging in humansMagnetic Resonance in Medicine, 2013
- Benefits and Harms of CT Screening for Lung CancerJAMA, 2012
- Comparison of lung T2* during free‐breathing at 1.5 T and 3.0 T with ultrashort echo time imagingMagnetic Resonance in Medicine, 2011
- Magnetic resonance imaging with ultrashort TE (UTE) PULSE sequences: Technical considerationsJournal of Magnetic Resonance Imaging, 2007