Stent deployment within a guide catheter aids removal of a fractured buddy wire
Authors:
- 24 May 2002
- journal article
- case report
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 56 (2) , 212-214
Abstract
We report a case of fractured buddy wire that was successfully removed by deploying a stent within the guide catheter, trapping the fractured segment of wire between stent and endoluminal surface of the guide catheter. This technique provides an alternative to either percutaneous snare or surgical intervention. Cathet Cardiovasc Intervent 2002;56:212–214.Keywords
This publication has 10 references indexed in Scilit:
- Broken guidewire fragment: A simplified retrieval techniqueCatheterization and Cardiovascular Interventions, 2000
- Dislodged stent: A simple retrieval techniqueCatheterization and Cardiovascular Interventions, 1999
- An unusual problem during surgical removal of a broken guidewireThe Annals of Thoracic Surgery, 1991
- Fractured and retained guide‐wire fragment during coronary angioplasty—unforeseen late sequelaeCatheterization and Cardiovascular Diagnosis, 1990
- Improved method for transcatheter retrieval of intracoronary detached angioplasty guidewire segmentsCatheterization and Cardiovascular Diagnosis, 1989
- Removal of guidewire fragmentThe Annals of Thoracic Surgery, 1989
- Surgical Removal of Guidewire Fragment Following Transluminal Coronary AngioplastyThe Annals of Thoracic Surgery, 1988
- Snare loop technique for removal of broken steerable PTCA wireCatheterization and Cardiovascular Diagnosis, 1987
- Successful percutaneous removal of retained broken coronary angioplasty guidewireCatheterization and Cardiovascular Diagnosis, 1986
- Intravascular retrieval of broken guide wire from the ascending aorta after percutaneous transluminal coronary angioplastyCatheterization and Cardiovascular Diagnosis, 1985