Determination of the Isotopic Composition of Ytterbium by MC-ICP-MS Using an Optimized Regression Model
Authors:
- 12 May 2022
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 94 (20) , 7200-7209
Abstract
In this study, we measure the absolute isotope ratios of ytterbium (Yb) by multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) using an optimized regression model for mass bias correction. A rhenium (Re) reference material (NIST SRM 3143), which has been characterized previously, is selected as a primary calibrator to calibrate the absolute Yb isotope ratios for three Yb materials (GSB, Alfa Yb, and GBW). The three-isotope plot for all collected data indicates that the results of Yb isotope ratios obtained are not affected by any polyatomic interferences and the mass-independent isotopic fractionation. Furthermore, the recalibrated Hf historical isotope ratios by using the absolute Yb isotopic composition obtained in this study for the isobaric interference correction on Hf isotopes are in agreement with the original historical values. This work has further demonstrated the applicability of the regression model for the calibrated measurements of absolute isotope ratios using MC-ICP-MS. The three mono-elemental Yb standard solutions are thus proposed as the reference materials for Yb isotope ratio measurements in environmental and geoscience applications.Keywords
Funding Information
- Ministry of Science and Technology of the People's Republic of China (2015CB856101, 2016YFC0600103)
- State Key Laboratory of Geological Processes and Mineral Resources (MSFGPMR01)
- Natural Science Foundation of Hubei Province (2020CFA045)
- National Natural Science Foundation of China (41803015, 41922021)
- Higher Education Discipline Innovation Project (BP0719022)
This publication has 58 references indexed in Scilit:
- Determination of mass-dependent variations in nickel isotope compositions using double spiking and MC-ICPMSJournal of Analytical Atomic Spectrometry, 2011
- Isotopic composition of Yb and the determination of Lu concentrations and Lu/Hf ratios by isotope dilution using MC‐ICPMSGeochemistry, Geophysics, Geosystems, 2004
- Accurate and High‐Precision Measurement of Lithium Isotopes in Two Reference Materials by MC‐ICP‐MSGeostandards and Geoanalytical Research, 2004
- Inter-laboratory and other errors in Pb isotope analyses investigated using a 207Pb–204Pb double spikeChemical Geology, 2000
- Precise analysis of copper and zinc isotopic compositions by plasma-source mass spectrometryChemical Geology, 1999
- Lead isotopic analyses of NIST Standard Reference Materials using multiple collector inductively coupled plasma mass spectrometry coupled with a modified external correction method for mass discrimination effectThe Analyst, 1996
- In situ hafnium isotope ratio analysis of zircon by inductively coupled plasma multiple collector mass spectrometryChemical Geology, 1995
- Table of the isotopic compositions of the elementsInternational Journal of Mass Spectrometry and Ion Processes, 1993
- Isotope fractionation laws: a test using calciumInternational Journal of Mass Spectrometry and Ion Processes, 1989
- Contributionàl'étude de la température dans les réacteurs fossiles d'Oklo par la mesure du rapport isotopique du lutétiumEarth and Planetary Science Letters, 1981