Abstract
A simple GC-MS method has been developed and validated for the direct determination of hexamethylenetetramine (HMT). The separation of HMT was performed using a MXT-1 column. The calibration curve was linear over the concentration range 0.1-25 μg/mL, with a good correlation coefficient (r =0.9996). The recoveries of HMT from foods spiked at 1, 5, and 10 μg/g ranged from 91.7% to 115.2%. Intra-day (n=5) and inter-day (n=5) precision were less than 7%. The limit of detection and the limit of quantification of the method were 0.05 and 0.15 μg/mL, respectively. The uncertainties associated with food matrix and calibration contributed most to the overall expanded uncertainty. The method validation data indicated that quantitative method could be applied to the direct determination of non-hydrolyzed HMT in foods.Keywords
This publication has 9 references indexed in Scilit:
- Determination of hexamethylenetetramine in foods by high-performance liquid chromatography (HPLC)Food Additives & Contaminants: Part A, 2014
- GC/MS: A practical user’s guide, 2nd edition Marvin McMasterJournal of the American Society for Mass Spectrometry, 2008
- Determination of methenamine, methenamine mandelate and methenamine hippurate in pharmaceutical preparations using ion-exchange HPLCJournal of Pharmaceutical and Biomedical Analysis, 2006
- Production of Hexamethylenetetramine in Photolyzed and Irradiated Interstellar Cometary Ice AnalogsThe Astrophysical Journal, 2001
- An integrated method for spectrum extraction and compound identification from gas chromatography/mass spectrometry dataJournal of the American Society for Mass Spectrometry, 1999
- A Novel Approach to Quantify the Amount of Formaldehyde Added to Milk in Grana Padano Cheese ProductionJournal of Food Science, 1989
- Separation and Quantitation of Methenamine in Urine by Ion-Pair ExtractionJournal of Pharmaceutical Sciences, 1986
- Gas Chromatographic Determination of Trace Amounts of Hexamethylenetetramine in Water Using an Iodine Charge-Transfer ComplexAnalytical Sciences, 1985
- Thermal Decomposition of Hexamethylene TetramineBulletin of the Chemical Society of Japan, 1968