Lower Tear Meniscus Height Measurements Using Keratography and Swept-Source Optical Coherence Tomography and Effect of Fluorescein Instillation Methods
Authors:
- 20 June 2019
- journal article
- research article
- Published by Taylor & Francis in Current Eye Research
- Vol. 44 (11) , 1203-1208
Abstract
Purpose: The first aim of this study was to compare inferior tear meniscus height (TMH) measured by keratography and swept-source optical coherence tomography (OCT). Effect of locations of fluorescein instillation and subjects’ subjective sensation, and the TMH results were also investigated. Methods: Forty-one eyes from 41 healthy adults had TMH randomly measured by keratography (K5M, Oculus, Germany) and swept-source OCT (Casia SS-1000, Tomey Corp, Japan) at baseline, followed by touching either the inferior bulbar or inferior palpebral conjunctiva, randomly assigned, with a moistened fluorescein strip. Two-way analysis of variance (ANOVA) was used to compare TMH obtained from the two devices and three measurement conditions. Subjective sensation of conjunctival touch was obtained through a 4-point scale with zero representing no sensation, and one to three representing ocular sensation in an ascending order. Results: Two-way ANOVA revealed no significant difference in TMH between keratography and swept-source OCT (Factor (devices): F = 0.579, p = 0.451) while there was significant difference among fluorescein instillations (Factor (conditions): F = 10.146, p < 0.001). Baseline TMH was around 0.25mm, and increased to around 0.28mm after instillation of fluorescein sodium. Ocular sensation was similar between the two conjunctival touch, and TMH was not significantly associated with ocular sensation. Conclusions: Keratography and swept-source OCT provided similar TMH results. Applying fluorescein sodium through moistened fluorescein strip could over-estimate TMH by around 0.03mm. Applying moistened fluorescein strip at bulbar or palpebral conjunctival led to similar subjective sensations and did not affect TMH measurement.Keywords
This publication has 31 references indexed in Scilit:
- New method for evaluation of early phase tear clearance by anterior segment optical coherence tomographyActa Ophthalmologica, 2013
- Dynamic Changes in the Lower Tear Meniscus After Instillation of Artificial TearsCornea, 2010
- Clinical assessment of the lower tear meniscus heightOphthalmic and Physiological Optics, 2009
- Prevalence of dry eye among adult Chinese in the Beijing Eye StudyEye, 2008
- Effect of Blinking on Tear DynamicsInvestigative Ophthalmology & Visual Science, 2007
- Quantitative Evaluation of Tear Meniscus Height From Fluorescein PhotographsCornea, 2007
- An introduction to analysis of variance (ANOVA) with special reference to data from clinical experiments in optometryOphthalmic and Physiological Optics, 2000
- The Cotton Thread Test: A Brief Review and a Clinical Study of Its Reliability on Hong Kong-ChineseOptometry and Vision Science, 1993
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Factors Affecting Tear Film Breakup in Normal EyesArchives of Ophthalmology (1950), 1973