A Low-Noise Analog Front End with Interstage Systematically Ambient Interference Cancellation for Pulse Oximeter
Open Access
- 9 March 2022
- journal article
- research article
- Published by MDPI AG in Electronics
- Vol. 11 (6) , 868
Abstract
This paper presents a low-noise analog front end (AFE) with interstage systematically ambient interference cancellation for a pulse oximeter, which is suitable for clinical oxygen saturation () detection with a low perfusion index. The fully differential implementation is adopted to have a better rejection performance of common mode interference and noise. Firstly, the proposed interstage systematically ambient interference cancellation is placed in the two gain stages to decrease low frequency noise in the bandwidth of interest (0.1–20 Hz), so that the larger signal-to-noise ratio (SNR) can be achieved to increase the detection accuracy of this system. Secondly, due to the additional gain stage compared with traditional implementation, the current-reuse OTA is employed to have better noise and power efficiency. Thirdly, the auto zero technique is utilized in the LED Driver to decrease the offset voltage and acquire a larger dynamic range (DR) in the low frequency bandwidth of interest. This PPG AFE chip is designed and fabricated in a 180 nm standard CMOS process. The receiver (RX) of this AFE consumes 220 from a 1.8 V supply, and the power consumption of the transmitter (TX) is 60 from a 3 V supply. The measurement results show that the input-referred noise current of 2.3 pA/sqrt(Hz) is achieved in RX and 110 dB peak DR is obtained in TX.
Keywords
Funding Information
- Natural Science Foundation of Chongqing, China (No. cstc2019jcyj-zdxmX0014)
This publication has 24 references indexed in Scilit:
- A 172 $\mu$W Compressively Sampled Photoplethysmographic (PPG) Readout ASIC With Heart Rate Estimation Directly From Compressively Sampled DataIEEE Transactions on Biomedical Circuits and Systems, 2017
- Multi-sensor fusion in body sensor networks: State-of-the-art and research challengesInformation Fusion, 2017
- Body Sensor Network for Mobile Health Monitoring, a Diagnosis and Anticipating SystemIEEE Sensors Journal, 2015
- A Low-Power, Dual-Wavelength Photoplethysmogram (PPG) SoC With Static and Time-Varying Interferer RemovalIEEE Transactions on Biomedical Circuits and Systems, 2014
- Enabling Effective Programming and Flexible Management of Efficient Body Sensor Network ApplicationsIEEE Transactions on Human-Machine Systems, 2012
- A Sub-mW Fully-Integrated Pulse Oximeter Front-EndIEEE Transactions on Biomedical Circuits and Systems, 2012
- Body Sensor Networks: A Holistic Approach From Silicon to UsersProceedings of the IEEE, 2011
- An Ultra-Low-Power Pulse Oximeter Implemented With an Energy-Efficient Transimpedance AmplifierIEEE Transactions on Biomedical Circuits and Systems, 2009
- Circuit techniques for reducing the effects of op-amp imperfections: autozeroing, correlated double sampling, and chopper stabilizationProceedings of the IEEE, 1996
- Switched-capacitor filtersIEEE Transactions on Power Electronics, 1990