Torsional properties of implant grade titanium
Authors:
- 1 March 1991
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 25 (3) , 281-293
Abstract
The purpose of this research was to evaluate the torsional strength and ductility of CP titanium in the as received condition, heat treated below the α → β transition temperature, and glass bead blasted. Results were compared to as‐received samples of implant quality 316L stainless steel. Little or no effect of any of the treatments was noted. The torsional strength of all titanium samples was comparable to 316L stainless and the torsional ductility of the titanium was significantly greater than 316L stainless (p <.01). These results are significantly different than those obtained when torsion testing bone screws of the same material. A hypothesis is proposed for the differences in these results.Keywords
This publication has 7 references indexed in Scilit:
- An analysis of the resistance of some dental post materials to failure in torsionClinical Materials, 1988
- Torsional elastic property measurements of selected orthodontic archwiresClinical Materials, 1987
- Effect of α phase morphology on mechanical properties of commercial purity titaniumMetal Science, 1983
- Mechanical properties of orthodontic wires in tension, bending, and torsionAmerican Journal of Orthodontics, 1982
- Comparison of the elastic properties of nickel-titanium and beta titanium arch wiresAmerican Journal of Orthodontics, 1982
- Deformation and unstable flow in hot torsion of Ti-6Ai-2Sn-4Zr-2Mo-0.1SiMetallurgical Transactions A, 1981
- Mechanical metallurgy.1961